[EN] English - K0394 Issue 3
Operation
11
Menu: Monitor
maximum/minimum
Use this function to monitor the maximum and minimum pressure.
It uses the specified scan rate (refer to menu operation, page 5).
Recommendation
To save battery power use an external power supply.
Maximum/minimum -
Set-up and use
Menu: Set function to
On
(refer to menu operation, page 5).To save
battery power use an external power supply.
When function is
On
, use steps
1
&
2
to show the
maximum/minimum since the last reset.
Step
3
resets the values for maximum/minimum, press and hold.
Menu: Monitor
a
pressure switch
Use this function to measure the performance of a pressure switch
(mechanical operation and hysteresis). It uses the specified scan
rate (refer to menu operation, page 5).
Recommendation
To save battery power use an external power supply.
Pressure switch input -
Set-up and use
1. Connect the DPI 104 with the applicable Table 3 accessories,
page 4:
Accessories B/C - refer to cover, Figure B2 (a).
Accessory E - refer to Figure 2, below/Table 6, page 8.
2. Menu: Set function to
On
(refer to menu operation, page 5).
Figure 2:
Example configuration - Switch input
These examples (Figure B2 (a)/, above) show the display when the
switch condition changes (open or closed). The analog dial and the
% indication continue to monitor the normal pressure.
1
maximum
2
minimum
3
donE
Reset
Pin 1 (+), 2 (GND)
12-24 Vdc
Pin 7 (Switch)
Pin 2 (GND)