18
K0394 Issue 3 - [EN] English
Maintenance
Cleaning
Inspecting
Repair
Recycling
Clean the case with a moist, lint-free cloth and a weak detergent.
Do not use solvents or abrasive materials.
Make sure that there is no damage to the threads and o-rings, free
of grit and other obstructions.
Return the instrument to the manufacturer or an approved service
agent for all repairs.
Do not dispose of this product as household waste. Use an
approved organisation that collects and/or recycles waste
electrical and electronic equipment.
Replace the batteries
To replace the batteries, refer to menu operation, page 5.
Note: When replacing batteries all the configuration options stay in
memory.
Restore the original
configuration
To restore the unit to the original factory configuration,
• Press and hold all four buttons until the display goes off (
≈
five
seconds).
• The unit will restart.
Menu operations shows the factory settings.
The lock code is reset to the factory code (000).
Calibration
Return the DPI 104 to the manufacturer or an approved service
agent for calibration.
Note: GE can provide a calibration service that is traceable to
international standards.
If using an alternative calibration facility, make sure that it uses
these standards.
Maintenance/Calibration