15

Getting Started

in a medical/hospital grade attachment plug, provided 
with the following cord tag: “Hospital Grade Plug." 
Grounding integrity can only be maintained when 
equipment is connected to an equivalent receptacle 
marked "Hospital Grade."

• Where the integrity of the external earth conductor in 

the installation or its arrangement is in doubt, the 
Monitor must be operated from its internal battery.

General Caution

• Do not touch either the pin of the DC input connector 

(3) or the terminals within the battery compartment (1) 
and the patient at the same time.

Содержание Dinamap PRO Series

Страница 1: ...1 DINAMAP PRO Series 100 400V2 Monitor Operation Manual...

Страница 2: ...2 DINAMAP PRO 400V2 Monitor...

Страница 3: ...P PRO Model 410 V2 refers to the second version of the product s design Reissues and Updates Changes occurring between issues are addressed through Change Information Sheets Addendums and replacement...

Страница 4: ...dure A general caution is a statement that alerts the user to the possibility of a problem with the device associated with its use or misuse Such problems include device malfunction device failure dam...

Страница 5: ...itor 23 Noninvasive Blood Pressure Determination 23 Procedures 27 Manual Mode 29 Auto Mode 29 Stat Mode 30 TURBO TEMP 31 Masimo SET SpO2 37 NELLCOR OxiMAXTM SpO2 46 Using the Menu System 55 Introducti...

Страница 6: ...sive Blood Pressure Determination 91 Appendix D 95 Reorder Codes 95 Appendix E 97 Warranty Service and Spare Parts 97 Repairs 98 Packing Instructions 98 Service Manuals 98 Appendix F 99 Maintenance 99...

Страница 7: ...lse and SpO2 PRO 400V2 BP Pulse Temp and SpO2 All of the main operations of the PRO Monitor are easy to use Please review the factory default settings and where applicable enter settings appropriate f...

Страница 8: ...e Monitor does not include any user replaceable fuses Refer servicing to qualified service personnel To reduce the risk of electric shock do not remove the cover or the back Refer servicing to a quali...

Страница 9: ...PRO Monitor Do not immerse the Monitor in water If the Monitor is splashed with water or becomes wet wipe it immediately with a dry cloth Do not gas sterilize or autoclave The PRO Monitor when used wi...

Страница 10: ...ent Type BF applied parts IPX1 degree of protection against ingress of water Sterilization Disinfection see Appendix F DINAMAP PRO MONITOR CLASSIFIED WITH RESPECT TO ELECTRIC SHOCK FIRE AND MECHANICAL...

Страница 11: ...Defibrillator proof type BF equipment Power ON OFF SpO2 START STOP BP AUTO BP Battery Power MAP Mean Arterial Pressure Temperature Beats Per Minute SILENCE External Communications Port Connector Exte...

Страница 12: ...atmospheric pressure range of 500 to 1060 hPa The DINAMAP PRO Monitor is protected against vertically falling drops of water and conforms to IEC 529 standard at level of IPX1 Vertically falling drops...

Страница 13: ...line input fuses are mounted internally and are replaceable only by qualified service personnel The remaining three fuses are auto resetable and mounted within the Monitor These fuses protect the low...

Страница 14: ...ced To run the Monitor on AC power it must be powered off and then on again Battery charging will take place as long as the Monitor remains connected to an external AC power source A battery that is f...

Страница 15: ...is connected to an equivalent receptacle marked Hospital Grade Where the integrity of the external earth conductor in the installation or its arrangement is in doubt the Monitor must be operated from...

Страница 16: ...cal Systems Information Technologies AC DC power converter ONLY 4 Inactive temperature cable storage Inactive temperature probe cable attaches here Models 200V2 and 400V2 5 Pole clamp Used to clamp mo...

Страница 17: ...splay 3 digit red LED indicates measured diastolic BP in mmHg 10 Temperature probe cover storage Box of probe covers stored here Models 200V2 and 400V2 11 Inactive temperature probe holster Extra temp...

Страница 18: ...t and select items in LCD menus if monitor is off pressing rotor will switch monitor on 22 Pulse BPM display 3 digit yellow LED shows pulse rate in beats per minute 23 SpO2 display 3 digit red LED ind...

Страница 19: ...nd automatically deflate the cuff When the Monitor is operating on the internal battery only battery life is enhanced by the use of the sleep mode However the PRO Monitor will not enter sleep mode if...

Страница 20: ...power the time and battery icon toggle and the BP and Printer modes Using the Printer Installing the Paper Turn the PRO Monitor so that the side is facing you While grasping the side of the Monitor fi...

Отзывы: