Revision A
DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual
7
2018554-001
Masimo unit). If the Monitor is equipped for Nellcor SpO
2
, this menu
allows the user to set the Response Mode and SatSeconds feature.
Alarms Button
This button accesses a dialog box to enter the alarms configuration
menu
.
Selecting
Yes
enters the menu. Selecting
No
returns the user
to the More
…
menu.
Select
Reset
to return all the alarm limits to the default settings. The
illustration above shows the default settings.
The user can adjust the alarm limits within the ranges specified in the
table below. The high limit must be at least one step higher than the
low limit.
Note: Predictive temperature has no alarm limit alarm.
Select
Save
to save the displayed settings and return to the
More…
menu. Select
Cancel
to abandon any changes and return to the
More…
menu.
Alarm Limits Table
Parameter
Range
Default
Systolic High
35 - 290
180
Systolic Low
30 - 285
30
Diastolic High
15 - 220
130
Diastolic Low
10 - 215
15
MAP High
25 - 260
140
MAP Low
20 - 255
50
BPM High
35 - 250
150
BPM Low
30 - 245
50
SpO2 High
21 - 100
Off
SpO2 Low
20 - 99
90
Содержание Dinamap PRO 100
Страница 1: ...PRO Series 100 400V2 Pre Service and Calibration Manual Models 110 410 ...
Страница 2: ......
Страница 4: ...DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001 ...
Страница 35: ...Revision A DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual 27 2018554 001 1 11 ERROR CODES ...
Страница 36: ...28 DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001 ...
Страница 37: ...Revision A DINAMAP PRO Series 100 400V2 Monitor Service Manual A 1 2018554 001 A Appendix ...
Страница 38: ...A 2 DINAMAP PRO Series 100 400V2 Monitor Service Manual Revision A 2018554 001 ...
Страница 41: ......