
Revision A
DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual
17
2018554-001
1.6.3 NIBP Calibration Check
This procedure verifies the linearity and calibration for both pressure
transducers (PT1 and PT2) over the range 0-250 mmHg.
To verify calibration it is necessary to have the following test
equipment:
- 2-tube NIBP hose, 12ft, p/n 107365
- DINAMAP Adult BP cuff p/n 2774 (or equivalent)
- Calibration Kit p/n 320246 with a manual inflation bulb
- NIST calibrated single-tube manometer
Figure 4-1
1. Set up the unit and calibration equipment as shown in Figure 4-1.
2. Power on the PRO Monitor and select
More...
3. Select the
Service
mode and input
2 2 1 3
.
4. From the
Service
menu, select the
NIBP
button.
5. From the
NIBP
menu, select the
Check
button.
Содержание Dinamap PRO 100
Страница 1: ...PRO Series 100 400V2 Pre Service and Calibration Manual Models 110 410 ...
Страница 2: ......
Страница 4: ...DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001 ...
Страница 35: ...Revision A DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual 27 2018554 001 1 11 ERROR CODES ...
Страница 36: ...28 DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001 ...
Страница 37: ...Revision A DINAMAP PRO Series 100 400V2 Monitor Service Manual A 1 2018554 001 A Appendix ...
Страница 38: ...A 2 DINAMAP PRO Series 100 400V2 Monitor Service Manual Revision A 2018554 001 ...
Страница 41: ......