Revision A
DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual
5
2018554-001
3. Clinician Mode requires a four-digit code. Turn the rotor to highlight
a number, and push to make a selection. Enter access code:
1 2 3 4.
1.4.2 Clinician Mode Main Menu
The Clinician Mode Main menu, pictured above, provides access to
the options normally required for routine use of the PRO Monitor. The
individual buttons on this menu are described briefly below.
Press Button
Selecting the
Press
button displays the target pressure dialog box,
pictured above, which sets the default target inflation pressure for a
BP cycle. The factory default is 150 mmHg for adults and 110 mmHg
for neonates. This is indicated by the “Auto” label at either end of the
adjustable range. Choosing “Retain Values” allows the user to preset
the number of minutes for which displayed values are retained on the
Monitor.
Selecting the
Info
button displays the most recent calibration dates of
the BP. Choose
OK
to return to the Service menu.
Содержание Dinamap PRO 100
Страница 1: ...PRO Series 100 400V2 Pre Service and Calibration Manual Models 110 410 ...
Страница 2: ......
Страница 4: ...DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001 ...
Страница 35: ...Revision A DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual 27 2018554 001 1 11 ERROR CODES ...
Страница 36: ...28 DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001 ...
Страница 37: ...Revision A DINAMAP PRO Series 100 400V2 Monitor Service Manual A 1 2018554 001 A Appendix ...
Страница 38: ...A 2 DINAMAP PRO Series 100 400V2 Monitor Service Manual Revision A 2018554 001 ...
Страница 41: ......