50
3.4.7 Reassembling the Monitor
Re-assembly is a direct reversal of disassembly, however additional care must be taken to ensure that all
cables and hoses are refitted correctly. The main points to note when reassembling are:
? ?
Pay particular attention to the wire dressing and ensure that all ferrite components are refitted
correctly.
? ?
When reassembling the chassis, remember to reconnect the valve control from the pump to the
power supply board.
? ?
When refitting the front panel, ensure that the copper earthing fingers of the front-panel engage
correctly inside the rear enclosure.
? ?
Ensure that the hoses to the front panel are not trapped or kinked while refitting the panel.
? ?
Keep your work area clean. It is quite easy for the plastic display panel cover to become
scratched if the front panel is pressed against a hard object such as a screw.
? ?
Be sure to re-calibrate and test the monitor as per section 3.4.1 and 3.4.6 before releasing
it for clinical use.
ZIF socket
Printer module
Printer board
8650AB