12
PREDICTIVE TEMPERATURE
?
0.6
o
C, 36.1
- 39.4
o
C ± 1
o
F, 97
- 103
o
F
ACCURACY :
UNSPECIFIED < 36.1
O
C, > 39.4
O
C
UNSPECIFIED < 97
O
F, > 103
O
F
DETERMINATION TIME:
30 seconds typical; 60 seconds maximum
SpO
2
(TS & S Model)
OXYGEN SATURATION
RANGE:
0 to 100%
SATURATION ACCURACY:
0% to 69%: unspecified
ADULT ACCURACY (70-100%)
Nellcor Puritan Bennett
™
Sensor
Accuracy
OXICLIQ-P Pediatric sensor
2.5 digits
OXIBAND Pediatric/infant sensor
3.0 digits
DURA-Y ear clip
3.5 digits
OXISENSOR II D-20 Pediatric sensor
3.5 digits
OXICLIQ-N neonatal/adult sensor
2.5 digits
REFLECTANCE sensor
3.5 digits
DURASENSOR adult
3.0 digits
OXIBAND adult/neonatal sensor
3.0 digits
DURA-Y multisite sensor
3.0 digits
OXISENSOR R-15 adult nasal sensor
3.5 digits
OXISENSOR II D-25 adult sensor
2.0 digits
OXICLIQ-A adult sensor
2.5 digits
OXISENSOR II N-25 neonatal/adult sensor
2.0 digits
OXISENSOR II I-20 infant sensor
2.0 digits
OXISENSOR II D-25L adult sensor, long cable
2.0 digits
Neonatal Accuracy:
When sensors are used on neonatal subjects as recommended, the specified accuracy range is increased
by
?
1 digit to account for the theoretical effect on oximeter measurements of fetal hemoglobin in neonatal
blood (e.g., N –25 accuracy on neonates is
?
3, rather than
?
2.)
Note: Refer to NELLCOR PURITAN BENNETT sensor specifications.
PULSE RATE DETERMINATION
:
20 BPM - 250 BPM +/- 3 beats
SATURATION PITCH INDICATOR:
Pitch changes with saturation.
Volume selectable from 0 (off) to 9.
WAVEFORMS:
Pulse Plethysmograph waveform on LCD gain compensated
SENSOR DISCONNECT /
The monitor will detect the attachment or disconnection
DISCONNECT FROM PATIENT:
of a sensor from the patient within 5 seconds.
PULSE DETECTION:
The monitor will detect a pulse or enter a no signal state within 15
seconds of being attached to the patient.