DigitalFlow™ GS868 Service Manual
21
Chapter 2. Error Codes and Screen Messages
2.16 E14: Totals Overflow
Problem:
The totalizers are unable to keep up with the total flow signals.
Cause:
The programmed units/pulse value is too small.
Action:
Select a larger number of units/pulse value.
2.17 E15: Equation Limit
Problem:
The meter cannot determine if the steam measurement is within 1% accuracy.
Cause:
The temperature or pressure reading is outside the accuracy specification of the GS868 steam equations.
(This error usually appears when the meter is running at a low temperature or the process has not come up
to speed.)
Action:
Adjust the pressure reading to within the range of 139.9685 kPa to 2242.9435 kPa, or the temperature
reading to within the range of 373.1500K to 810.9278K.
Содержание DigitalFlow GS868
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 10: ...Contents viii DigitalFlow GS868 Service Manual...
Страница 26: ...Chapter 1 Calibration DigitalFlow GS868 Service Manual 16...
Страница 34: ...Chapter 2 Error Codes and Screen Messages 24 DigitalFlow GS868 Service Manual no content intended for this page...
Страница 40: ...Chapter 3 Diagnostics 30 DigitalFlow GS868 Service Manual no content intended for this page...
Страница 54: ...Appendix A Service Record 44 DigitalFlow GS868 Service Manual no content intended for this page...
Страница 62: ...Appendix B Optional Enclosures 52 DigitalFlow GS868 Service Manual no content intended for this page...
Страница 68: ...Warranty 58 DigitalFlow GS868 Service Manual no content intended for this page...
Страница 72: ......
Страница 73: ......