Chapter 1. Calibration
14
DigitalFlow™ GS868 Service Manual
1.8.3 Entering Pulse Number
5.
Enter the number of pulses desired (between 1 and 10,000) and press the
[ENT]
key. That number of pulses will then
be output at the specified frequency.
6.
Repeat Steps 3, 4 and 5 to test all four of the frequency/totalizer outputs. If any of the outputs fails to pass the test,
contact the factory for assistance.
Procedure Options
You have completed testing the totalizer/frequency outputs. Do one of the following:
•
To calibrate/test additional slots, press
[F4] (EXIT)
and proceed to the appropriate section.
•
To leave the Calibration Menu, press
[F4] (EXIT)
twice.
Содержание DigitalFlow GS868
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 10: ...Contents viii DigitalFlow GS868 Service Manual...
Страница 26: ...Chapter 1 Calibration DigitalFlow GS868 Service Manual 16...
Страница 34: ...Chapter 2 Error Codes and Screen Messages 24 DigitalFlow GS868 Service Manual no content intended for this page...
Страница 40: ...Chapter 3 Diagnostics 30 DigitalFlow GS868 Service Manual no content intended for this page...
Страница 54: ...Appendix A Service Record 44 DigitalFlow GS868 Service Manual no content intended for this page...
Страница 62: ...Appendix B Optional Enclosures 52 DigitalFlow GS868 Service Manual no content intended for this page...
Страница 68: ...Warranty 58 DigitalFlow GS868 Service Manual no content intended for this page...
Страница 72: ......
Страница 73: ......