Chapter 1. Calibration
8
DigitalFlow™ GS868 Service Manual
1.5.5 20 mA Option = [F4]
1.
Connect the 20 mA current source to the currently selected analog input, as shown in Figure 3 on page 6.
2.
Press
[F1]
to store the current 20 mA value or press
[F2]
to cancel the entry. In either case, the
ANALOG INPUT
prompt
reappears.
3.
Go to
Procedure Options
below.
Procedure Options
You have completed calibration of the analog output. Do one of the following:
•
To calibrate another input, return to Step 3 in
Accessing the Calibration Menu
on the previous page.
•
To calibrate/test additional slots, press
[F4] (EXIT)
and proceed to the appropriate section.
•
To leave the Calibration Menu, press
[F4] (EXIT)
twice.
Содержание DigitalFlow GS868
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 10: ...Contents viii DigitalFlow GS868 Service Manual...
Страница 26: ...Chapter 1 Calibration DigitalFlow GS868 Service Manual 16...
Страница 34: ...Chapter 2 Error Codes and Screen Messages 24 DigitalFlow GS868 Service Manual no content intended for this page...
Страница 40: ...Chapter 3 Diagnostics 30 DigitalFlow GS868 Service Manual no content intended for this page...
Страница 54: ...Appendix A Service Record 44 DigitalFlow GS868 Service Manual no content intended for this page...
Страница 62: ...Appendix B Optional Enclosures 52 DigitalFlow GS868 Service Manual no content intended for this page...
Страница 68: ...Warranty 58 DigitalFlow GS868 Service Manual no content intended for this page...
Страница 72: ......
Страница 73: ......