March 2008
1-38
Programming Site Data
Setting up Alarms
1.
Use the
[F1]-[F3]
keys to set up alarm relays A, B, or C,
respectively.
Note:
The set up of alarm A is used as an example. Identical
procedures would be used to set up the other alarms.
2.
Press
[F1] = OFF
to disable Alarm A and return to the previous
prompt, or press
[F2] = HIGH, [F3] = LOW or [F4]
= FAULT
to specify
the type for Alarm A.
3.
Press
[F1] = NO
for standard operation or
[F2] = YES
for
Failsafe
operation.
4.
Use the [
←]
and [
→]
and
[F1]-[F4]
keys to select a
Channel
option
for the data source.
5.
Do one of the following:
•
If you selected
FAULT
,
proceed to step 6.
•
If you selected
HIGH
or
LOW
,
complete the steps below.
a.
Use the [
←]
and [
→]
and
[F1]-[F4]
keys to specify the desired
Output Measurement
parameter.
b.
Enter a value for the trigger point of the alarm and press
[ENT]
.
6.
To select the
Fault Type
that will trigger the alarm, press
[F1] =
FLOW, [F2] = OTHER
(non-flow), or
[F3]= ALL
.
7.
Press
[F1]
-
[F3]
to set up another alarm or press
[EXIT]
to select
another slot for set up.
Содержание DigitalFlow GS868
Страница 7: ...Chapter 1 ...
Страница 59: ...Chapter 2 ...
Страница 77: ...Chapter 3 ...
Страница 90: ...Chapter 4 ...
Страница 104: ...Chapter 5 ...
Страница 105: ...Clearing Data Introduction 5 1 Resetting Totals 5 2 Deleting Site Files 5 3 Deleting Log Files 5 4 ...
Страница 110: ...Chapter 6 ...
Страница 122: ...Appendix A ...
Страница 131: ...Appendix B ...
Страница 132: ...Data Records Option Cards Installed B 1 Initial Setup Data B 2 ...
Страница 136: ...Appendix C ...
Страница 156: ...Appendix D ...