Programming Site Data
1-25
March 2008
Setting Response Time
Use this option to specify the number of readings that occur before
the meter will respond to a step change in flow rate. In general, the
smaller the number of readings, the less steady the display will
appear.
•
Use the [
→]
and
[F1]-[F4]
keys to select the desired option. The
available response time options are
1, 2, 5, 10, 30, 60 and STATS
. For
best results, select
STATS
(statistics), as this increases the response
time under steady flow conditions while still allowing a rapid
response to changes in flow rate.
The GS868 program returns to the
SETUP
window. Remember to
record all programmed data in Appendix B,
Data Records
.
Initializing the System
Use this option to initialize (reset) all of the parameters within the
SETUP
menu to their default values.
•
Press
[F1] = NO
to keep the current values or press
[F2] = YES
to reset
all values to their default settings.
Содержание DigitalFlow GS868
Страница 7: ...Chapter 1 ...
Страница 59: ...Chapter 2 ...
Страница 77: ...Chapter 3 ...
Страница 90: ...Chapter 4 ...
Страница 104: ...Chapter 5 ...
Страница 105: ...Clearing Data Introduction 5 1 Resetting Totals 5 2 Deleting Site Files 5 3 Deleting Log Files 5 4 ...
Страница 110: ...Chapter 6 ...
Страница 122: ...Appendix A ...
Страница 131: ...Appendix B ...
Страница 132: ...Data Records Option Cards Installed B 1 Initial Setup Data B 2 ...
Страница 136: ...Appendix C ...
Страница 156: ...Appendix D ...