Programming Site Data
1-21
March 2008
Setting Transducer Signal
Limits (cont.)
5.
Press
[ENT]
to accept the current
Velocity High Limit
value or enter
a new value and press
[ENT]
.
The default value for this parameter is 275.0 ft/sec (85 m/sec) and
values from –500 to 500 ft/sec (–150 to 150 m/sec) are acceptable.
The
E3: VELOCITY RANGE
error message appears when the calculated
fluid velocity exceeds the programmed
VELOCITY HIGH LIMIT
value.
See the
Service Manual
for a discussion of error codes.
6.
Press
[ENT]
to accept the current
Acceleration
value or enter a new
value and press
[ENT]
.
The default value for this parameter is 100.0 ft/sec
2
(30.48 m/sec
2
)
and values from 0 to 250 ft/sec
2
(0 to 76 m/sec
2
) are acceptable. The
E6: CYCLE SKIP
error message appears when the calculated fluid
velocity changes by more than the programmed
ACCELERATION LIMIT
value from one reading to the next. See the
Service Manual
for a
discussion of error codes.
7.
Press
[ENT]
to accept the current
Amp. Discrim Low
value or enter
a new value and press
[ENT]
.
IMPORTANT:
Consult the factory before performing this step.
The amplitude discriminator measures the size of the transducer
signal received by the Model GS868. The default value for this
parameter is 14 and values from 0 to 100 are acceptable. The
E5:
AMPLITUDE
error message appears when the amplitude discriminator
falls below the programmed
AMP. DISCRIM LOW
value. See the
Service
Manual
for a discussion of error codes.
8.
Press
[ENT]
to accept the current
Amp. Discrim High
value or enter
a new value and press
[ENT]
.
IMPORTANT:
Consult the factory before performing this step.
The amplitude discriminator measures the size of the transducer
signal received by the Model GS868. The default value for this
parameter is 34 and values from 0 to 100 are acceptable. The
E5:
AMPLITUDE
error message appears when the amplitude discriminator
exceeds the programmed
AMP. DISCRIM HIGH
value. See the
Service
Manual
for a discussion of error codes.
9.
Press
[ENT]
to accept the current
Delta T Offset
value or enter a
new value and press
[ENT]
.
IMPORTANT:
Consult the factory before performing this step.
Содержание DigitalFlow GS868
Страница 7: ...Chapter 1 ...
Страница 59: ...Chapter 2 ...
Страница 77: ...Chapter 3 ...
Страница 90: ...Chapter 4 ...
Страница 104: ...Chapter 5 ...
Страница 105: ...Clearing Data Introduction 5 1 Resetting Totals 5 2 Deleting Site Files 5 3 Deleting Log Files 5 4 ...
Страница 110: ...Chapter 6 ...
Страница 122: ...Appendix A ...
Страница 131: ...Appendix B ...
Страница 132: ...Data Records Option Cards Installed B 1 Initial Setup Data B 2 ...
Страница 136: ...Appendix C ...
Страница 156: ...Appendix D ...