Programming Site Data
1-19
March 2008
Entering Setup Data
The signal limits and response times for the Model GS868 are
specified via the
SETUP
submenu. While following the programming
instructions, refer to the menu map in Figure A-2 on page A-2 in
Appendix A,
Menu Maps
. The following three submenus are included
in this section:
•
SIGNL
- set the parameters related to the transducer signal
•
AVRG
- specify the response of the meter to step changes
•
INIT
- initialize all parameters to default values.
•
ADVAN
- setting parameters for mass flow and multiple K-factors.
Enter the
SETUP
submenu by pressing [
→]
and
[F1]
at the
Channel
PROGRAM
prompt and complete the following steps:
Proceed to the appropriate section to program the option selection
made at the above prompt. Remember to record all programmed data
in Appendix B,
Data Records
.
Chan 1 SETUP
Press
[F1]-[F3]
to select the
desired
SETUP
option.
Channel PROGRAM
SET UP
SET UP
last selection appears here
SIGNL
AVRG
INIT
ADVAN
Содержание DigitalFlow GS868
Страница 7: ...Chapter 1 ...
Страница 59: ...Chapter 2 ...
Страница 77: ...Chapter 3 ...
Страница 90: ...Chapter 4 ...
Страница 104: ...Chapter 5 ...
Страница 105: ...Clearing Data Introduction 5 1 Resetting Totals 5 2 Deleting Site Files 5 3 Deleting Log Files 5 4 ...
Страница 110: ...Chapter 6 ...
Страница 122: ...Appendix A ...
Страница 131: ...Appendix B ...
Страница 132: ...Data Records Option Cards Installed B 1 Initial Setup Data B 2 ...
Страница 136: ...Appendix C ...
Страница 156: ...Appendix D ...