8-2
Dash™ 3000/4000/5000
2000966-386D
Monitoring ECG
Introduction
Overview
ECG is the primary patient monitoring parameter. It measures heart rate, analyzes
arrhythmia, detects pacemaker function and detects myocardial ischemia.
A parameter window and waveform display when a patient cable is connected to the
monitor. The display updates every two seconds. The The parameter window varies
by software package and the control settings.
655A
When the quality of an electrode signal degrades to an inadequate monitoring level, a
lead failure message displays and monitoring automatically switches to another lead,
as described in the following table.
1
QRS indicator
2
Heart rate
3
Paced beat detection indicator (Pacemaker detection enabled)
4
PVC count (full arrhythmia only)
5
ST measurement point
6
ST analysis data (ST analysis only)
Message
New Lead Monitored
RA FAIL
Lead III
LL FAIL
Lead I
LA FAIL
Lead II
V FAIL
Lead II
LEADS FAILED
No waveforms displayed when RL or all leads fail.
3
4
5
6
2
1
Содержание Dash 3000
Страница 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Страница 12: ...x Dash 3000 4000 5000 2000966 386D ...
Страница 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Страница 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Страница 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Страница 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Страница 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Страница 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Страница 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Страница 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Страница 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Страница 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Страница 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Страница 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Страница 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Страница 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Страница 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Страница 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Страница 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Страница 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Страница 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Страница 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Страница 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Страница 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Страница 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Страница 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Страница 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Страница 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Страница 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Страница 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Страница 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Страница 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Страница 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Страница 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Страница 269: ......