Revision C
250cx Series Maternal/Fetal Monitor
D-9
2036946-001
Frequently Asked Questions:
How do I enable or change alarm volume
settings for Fetal Heart Rate 2?
1. Rotate the Trim Knob to highlight the legend for FHR2. (This legend is top left on the
display, and may display as INOP or US2.
2. Once the FHR2 legend is highlighted, press the Trim Knob. The display changes to
show the <MODE> Setup screen, where mode is the current legend.
3. Now rotate the Trim Knob to highlight the FHR2 Audio Alarms: setting. This is at the
bottom of the flat panel display, slightly above the menu bar, on the left half of the
screen. The setting is either On or Off. If the setting is Off, you must enable it.
Proceed to (Step 4) below. If the setting is already On, and you simply wish to
change the Volume: setting, proceed to (Step 7) below.
4. Once the Audio Alarms: setting is highlighted, press the Trim Knob again. The
current setting displays in blinking inverse video.
5. Now rotate the Trim Knob to change the current setting to On.
6. Once you set the desired alarm value, press the Trim Knob to confirm your selection.
The current value setting stops blinking.
7. Now rotate the Trim Knob to select (highlight) the Volume: setting located
immediately to the right. The setting is in the range of 1 to 9.
8. Once the Volume: setting is highlighted, press the Trim Knob again. The current
setting displays in blinking inverse video, and an audio tone, indicative of the alarm
volume, will be emitted from the speaker.
9. Now rotate the Trim Knob to change the current setting as desired. With each setting
change, an audio tone, indicative of the alarm volume, will be emitted from the
speaker.
10. Once you set the desired alarm value, press the Trim Knob to confirm your
selection. The current value setting stops blinking. A final audio tone, indicative of the
alarm volume, is emitted from the speaker.
11. Now rotate the Trim Knob to select (highlight) the Exit item on the bottom menu.
12. Press the Trim Knob again to return to the main monitoring display screen.
How do I enable or change alarm volume
settings for Non-Invasive Blood Pressure?
1. Rotate the Trim Knob to highlight the legend for NIBP. (This legend is slightly above
center, on the left side of the display.)
2. Once the NIBP legend is highlighted, press the Trim Knob. The display changes to
show the NIBP Setup screen.
3. Now rotate the Trim Knob to select (highlight) the Volume: setting which is located
immediately to the right. The setting is in the range of 1 to 9.
4. Once the Volume: setting is highlighted, press the Trim Knob again. The current
setting displays in blinking inverse video, and an audio tone, indicative of the alarm
volume, is emitted from the speaker.
5. Now rotate the Trim Knob to change the current setting as desired. Each time you
change a setting, an audio tone, indicative of the alarm volume, will be emitted from
the speaker.
6. Once you set the desired alarm value, press the Trim Knob to confirm your selection.
The current value setting stops blinking. A final audio tone, indicative of the alarm
volume, is emitted from the speaker.
7. Now rotate the Trim Knob to select (highlight) the Exit item on the bottom menu.
8. Press the Trim Knob again to return to the main monitoring display screen.
Question
Answer
Содержание Corometrics 250cx Series
Страница 2: ......
Страница 17: ...Revision C 250cx Series Maternal Fetal Monitor 1 1 2036946 001 1 Safety...
Страница 18: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety For your notes...
Страница 28: ...1 12 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety Equipment Symbols...
Страница 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036946 001 2 Introduction...
Страница 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Introduction For your notes...
Страница 35: ...Revision C 250cx Series Maternal Fetal Monitor 3 1 2036946 001 3 Controls Indicators and Connectors...
Страница 53: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036946 001 4 Setup Procedures...
Страница 54: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Setup Procedures For your notes...
Страница 65: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036946 001 5 Fetal Heart Rate Monitoring...
Страница 66: ...5 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Fetal Heart Rate Monitoring For your notes...
Страница 77: ...Revision C 250cx Series Maternal Fetal Monitor 6 1 2036946 001 6 Uterine Activity Monitoring...
Страница 78: ...6 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Uterine Activity Monitoring For your notes...
Страница 83: ...Revision C 250cx Series Maternal Fetal Monitor 7 1 2036946 001 7 Maternal Heart Pulse Rate Monitoring...
Страница 105: ...Revision C 250cx Series Maternal Fetal Monitor 9 1 2036946 001 9 Maternal Pulse Oximetry Monitoring...
Страница 106: ...9 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Maternal Pulse Oximetry Monitoring For your notes...
Страница 117: ...Revision C 250cx Series Maternal Fetal Monitor 10 1 2036946 001 10 Alarms...
Страница 118: ...10 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Alarms For your notes...
Страница 127: ...Revision C 250cx Series Maternal Fetal Monitor 11 1 2036946 001 11 Recorder Modes...
Страница 128: ...11 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Recorder Modes For your notes...
Страница 147: ...Revision C 250cx Series Maternal Fetal Monitor 12 1 2036946 001 12 Maternal Vital Signs History...
Страница 148: ...12 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Maternal Vital Signs History For your notes...
Страница 153: ...Revision C 250cx Series Maternal Fetal Monitor 13 1 2036946 001 13 Heartbeat Coincidence...
Страница 154: ...13 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Heartbeat Coincidence For your notes...
Страница 159: ...Revision C 250cx Series Maternal Fetal Monitor 14 1 2036946 001 14 Waveforms...
Страница 160: ...14 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Waveforms For your notes...
Страница 165: ...Revision C 250cx Series Maternal Fetal Monitor 15 1 2036946 001 15 Maintenance...
Страница 166: ...15 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Maintenance For your notes...
Страница 172: ...15 8 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Maintenance Disposal of Product Waste...
Страница 174: ...16 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Troubleshooting For your notes...
Страница 182: ...17 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Technical Specifications For your notes...
Страница 192: ...17 12 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Technical Specifications Strip Chart Recorder...
Страница 194: ...18 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Supplies Accessories For your notes...
Страница 200: ...A 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Factory Defaults For your notes...
Страница 206: ...A 8 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Factory Defaults Table of Defaults...
Страница 208: ...B 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Fetal Movement Detection For your notes...
Страница 212: ...B 6 250cx Series Maternal Fetal Monitor Revision C 2036946 001...
Страница 214: ...C 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Spectra Alerts For your notes...
Страница 234: ...C 22 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Spectra Alerts Alert Parameters...
Страница 235: ...Revision C 250cx Series Maternal Fetal Monitor D 1 2036946 001 Frequently Asked Questions D Frequently Asked Questions...
Страница 236: ...D 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Frequently Asked Questions...
Страница 249: ......