Revision C
250cx Series Maternal/Fetal Monitor
D-5
2036946-001
Frequently Asked Questions:
How do I enable the MHR/P, Maternal
Heart Rate Pulse, trend recorder tracing?
1. Rotate the Trim Knob to highlight the legend for MHR/P. (This legend is located
approximately in the center of the display, and may indicate MECG, Pulse or INOP,
according on the settings that are currently enabled.
2. Once the MHR/P legend is highlighted, press the Trim Knob. The display changes to
show the MHR/P Setup screen.
3. Now rotate the Trim Knob to highlight the HR/PR: setting. (It should read Off.)
4. Once the HR/PR: setting is highlighted, press the Trim Knob again. The current
setting displays in blinking inverse video.
5. Now rotate the Trim Knob to change the current setting. Select a setting of On (to
enable MHR/P trace) or Off.
6. Once you set the desired trace setting, press the Trim Knob to confirm your
selection. The current setting stops blinking.
7. Now rotate the Trim Knob to select (highlight) the Exit item on the bottom menu.
8. Press the Trim Knob again to return to the main monitoring display screen.
How do I change the source parameter for
MHR/P, Maternal Heart Rate Pulse?
1. Rotate the Trim Knob to highlight the legend for MHR/P. This legend is located
approximately in the center of the display, and may indicate MECG, Pulse or INOP
according to the settings that are currently enabled.
2. Once the MHR/P legend is highlighted, press the Trim Knob. The display changes to
show the MHR/P Setup screen.
3. Now rotate the Trim Knob to highlight the Source: setting. (It will read Auto, MSpO
2
,
MECG.)
4. Once the Source: setting is highlighted, press the Trim Knob again. The current
setting displays in blinking inverse video.
5. Now rotate the Trim Knob to change the current setting. Select from Auto, MSpO
2
,
MECG. If you select Auto, the pulse value parameter is automatically selected
according to the parameters that are currently enabled with precedence, highest to
lowest, in the following order: MECG, MSpO
2
.
6. Once you set the desired source parameter, press the Trim Knob to confirm your
selection. The current value setting stops blinking.
7. Now rotate the Trim Knob to select (highlight) the Exit item on the bottom menu.
8. Press the Trim Knob again to return to the main monitoring display screen.
How do I enable the MSpO
2
, Maternal
Blood Oxygen Saturation, trend recorder
tracing?
1. Rotate the Trim Knob to highlight the legend for MSpO
2
. (This legend is located
above the center, on the right side of the display.)
2. Once the MSpO
2
legend is highlighted, press the Trim Knob. The display changes to
show the MSpO
2
Setup screen.
3. Now rotate the Trim Knob to highlight the %O
2
Trace: setting. (It should read Off.)
4. Once the %O
2
Trace: setting is highlighted, press the Trim Knob again. The current
setting displays in blinking inverse video.
5. Now rotate the Trim Knob to change the current setting. Select a setting of On (to
enable MSpO
2
trace) or Off.
6. Once you set the desired trace setting, press the Trim Knob to confirm your
selection. The current setting stops blinking.
7. Now rotate the Trim Knob to select (highlight) the Exit item on the bottom menu.
8. Press the Trim Knob again to return to the main monitoring display screen.
Question
Answer
Содержание Corometrics 250cx Series
Страница 2: ......
Страница 17: ...Revision C 250cx Series Maternal Fetal Monitor 1 1 2036946 001 1 Safety...
Страница 18: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety For your notes...
Страница 28: ...1 12 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety Equipment Symbols...
Страница 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036946 001 2 Introduction...
Страница 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Introduction For your notes...
Страница 35: ...Revision C 250cx Series Maternal Fetal Monitor 3 1 2036946 001 3 Controls Indicators and Connectors...
Страница 53: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036946 001 4 Setup Procedures...
Страница 54: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Setup Procedures For your notes...
Страница 65: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036946 001 5 Fetal Heart Rate Monitoring...
Страница 66: ...5 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Fetal Heart Rate Monitoring For your notes...
Страница 77: ...Revision C 250cx Series Maternal Fetal Monitor 6 1 2036946 001 6 Uterine Activity Monitoring...
Страница 78: ...6 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Uterine Activity Monitoring For your notes...
Страница 83: ...Revision C 250cx Series Maternal Fetal Monitor 7 1 2036946 001 7 Maternal Heart Pulse Rate Monitoring...
Страница 105: ...Revision C 250cx Series Maternal Fetal Monitor 9 1 2036946 001 9 Maternal Pulse Oximetry Monitoring...
Страница 106: ...9 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Maternal Pulse Oximetry Monitoring For your notes...
Страница 117: ...Revision C 250cx Series Maternal Fetal Monitor 10 1 2036946 001 10 Alarms...
Страница 118: ...10 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Alarms For your notes...
Страница 127: ...Revision C 250cx Series Maternal Fetal Monitor 11 1 2036946 001 11 Recorder Modes...
Страница 128: ...11 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Recorder Modes For your notes...
Страница 147: ...Revision C 250cx Series Maternal Fetal Monitor 12 1 2036946 001 12 Maternal Vital Signs History...
Страница 148: ...12 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Maternal Vital Signs History For your notes...
Страница 153: ...Revision C 250cx Series Maternal Fetal Monitor 13 1 2036946 001 13 Heartbeat Coincidence...
Страница 154: ...13 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Heartbeat Coincidence For your notes...
Страница 159: ...Revision C 250cx Series Maternal Fetal Monitor 14 1 2036946 001 14 Waveforms...
Страница 160: ...14 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Waveforms For your notes...
Страница 165: ...Revision C 250cx Series Maternal Fetal Monitor 15 1 2036946 001 15 Maintenance...
Страница 166: ...15 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Maintenance For your notes...
Страница 172: ...15 8 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Maintenance Disposal of Product Waste...
Страница 174: ...16 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Troubleshooting For your notes...
Страница 182: ...17 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Technical Specifications For your notes...
Страница 192: ...17 12 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Technical Specifications Strip Chart Recorder...
Страница 194: ...18 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Supplies Accessories For your notes...
Страница 200: ...A 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Factory Defaults For your notes...
Страница 206: ...A 8 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Factory Defaults Table of Defaults...
Страница 208: ...B 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Fetal Movement Detection For your notes...
Страница 212: ...B 6 250cx Series Maternal Fetal Monitor Revision C 2036946 001...
Страница 214: ...C 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Spectra Alerts For your notes...
Страница 234: ...C 22 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Spectra Alerts Alert Parameters...
Страница 235: ...Revision C 250cx Series Maternal Fetal Monitor D 1 2036946 001 Frequently Asked Questions D Frequently Asked Questions...
Страница 236: ...D 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Frequently Asked Questions...
Страница 249: ......