Programming Site Data
2-5
September 2004
Special Transducers
1.
For a special transducer, press [
X
] to move the cursor to
STD
and
press
[
S
]
until
SPEC
appears and press
[ENTER]
.
2.
Assign a number between 90 and 99 to the special transducer.
Enter the number by using the arrow keys. Press [
W
]
and [
X
] to
move the cursor to the desired location and press
[
S
]
and
[
T
]
to
increase or decrease the number. Then press
[ENTER]
.
3.
Press [
W
]
and [
X
] to select the wedge type (supplied by GE) and
press
[ENTER]
.
4.
Press [
W
]
and [
X
] to select the transducer frequency (supplied by
GE) and press
[ENTER]
.
5.
Enter the special transducer time delay value (supplied by GE) by
using the arrow keys. Press [
W
]
and [
X
] to move the cursor to
desired location and press
[
S
]
and
[
T
]
to increase or decrease
the number and press
[ENTER]
.
Special Clamp-On
Transducers
The following two prompts appear only if special clamp-on
transducers are being used. If special wetted transducers are being
used, proceed to
Pipe Outside Diameter
on page 2-7.
1.
Enter the wedge angle (supplied by GE) of the transducer by using
the arrow keys. Press [
W
]
and [
X
] to move the cursor to the
desired location and press
[
S
]
and
[
T
]
to increase or decrease
the number. Then press
[ENTER]
.
2.
Enter the wedge sound speed (supplied by GE) of the transducer
by using the arrow keys. Press [
W
]
and [
X
] to move the cursor to
the desired location and press
[
S
]
and
[
T
]
to increase or
decrease the number. Then press
[ENTER]
and proceed to
Pipe
Material
on the next page.