September 2004
2-2
Programming Site Data
Accessing the ACTIV
Submenu
1.
Press
[ESC]
,
[ENTER]
,
[ESC]
in sequence to enter the
user program
.
2.
Press [
X
] until
PROG
appears and press
[ENTER]
.
3.
Press [
X
] until the desired channel appears. Press
[ENTER]
at
CH1
or
CH2
.
4.
Press [
X
] until
ACTIV
appears and press
[ENTER]
.
5.
Press [
X
] until
TRANS
appears and press
[ENTER]
. Selecting
TRANS
activates the channel.
6.
If you have a 2-channel unit, you can also disable a channel.
However, if you select
OFF
for a channel, you will not be able to
program that channel.
After completing the above step, the
user program
returns to the
Channel PROGRAM
menu. To continue entering your initial setup
data, proceed to the next section.
Entering System Data
[CHx-SYSTM Menu]
The
CHx-SYSTM
submenu (see
Figure A-1
on page A-1) is used to
enter system parameters for the individual channels. In situations
where the channels operate independently, the AT868 uses the system
parameters in this menu. When channels are averaged together, the
AT868 uses the parameters entered in the
GLOBL-SYSTM
submenu.
Accessing the CHx-SYSTM
Submenu
1.
Press
[ESC]
,
[ENTER]
,
[ESC]
in sequence to enter the
user program
.
2.
Press [
X
] until
PROG
appears and press
[ENTER]
.
3.
Press [
X
] until the desired channel appears. Press
[ENTER]
at
CH1
or
CH2
.
4.
Press [
X
] until
SYSTM
appears and press
[ENTER]
.
5.
Enter the desired
CHANNEL LABEL
(up to 5 characters) by using
the arrow keys. Press [
W
]
and [
X
] to move the cursor to the
desired location and press
[
S
]
and
[
T
]
to change a letter. Then
press
[ENTER]
.
6.
Use the arrow keys to enter the desired
SITE MESSAGE
(up to 16
characters) and press
[ENTER]
. (For a 2-channel meter, this prompt
is called
CHANNEL MESSAGE
.)