The image appears misty
Then...
If...
Turn off the instrument with the On/Off button on the front
panel and wait until the ambient environments meets the
specifications
Dew is condensing in
the optical system
Unevenness appears on the
image
Then...
If...
Turn off the instrument and contact your GE representa-
tive
The LED is deteriorated
from age
Turn off the instrument and contact your GE representa-
tive
A lamp in the UV transil-
luminator is broken
Noise appears in an image
exposed using the UV Trans tray
Then...
If...
The UV Trans Tray is contaminated. Clean the tray.
The UV Trans Tray is de-
teriorated from age
Artefacts appear in a
Colorimetric Epi image
Then...
If...
Rotate the gel sample 90° positioning the gel so that the
lanes of the gel are parallel to the long side of the tray.
The correct tray is used
Amersham Imager 680 Operating Instructions 29271308 AB
239
8 Troubleshooting
8.2 Problems with image quality