Fluorescence Trans-UV
Capturing with the Fluorescence image capturing method option Trans-UV is performed
using the UV Trans Tray (transparent tray) and a gel sheet.
Fluorescence Trans-UV sample preparation
Follow the instructions below to prepare the sample.
Action
Step
Cut out a gel sheet slightly larger than the sample size.
1
Place the gel sheet on the UV Trans Tray (transparent tray).
2
Place the sample on the gel sheet.
3
Proceed with the capturing.
Note:
After use, wash the gel sheet with mild detergent, rinse with water then dry
well. A gel sheet can be reused about 20 times.
4
Follow the instructions below to perform fluorescence capture using Fluorescence UV.
Action
Step
In the Capture tab, select Fluorescence. The options available for Fluores-
cence capturing are displayed to the right on the screen.
1
Choose the UV.
2
Amersham Imager 680 Operating Instructions 29271308 AB
159
5 Operation
5.5 Capture an image
5.5.4 Fluorescence capture