20
Valve INV-907 Instructions 56-8101-01 Edition AE
6 Reference information
Physical data
Internal volume
Pos 1 LOAD
Port
1–7
9
μl
Port 2–3
5 μl
Port 4–6
8 μl
Pos 2 INJECT
Port
1–2
7
μl
Port 3–5
6 μl
Port
6–7
9
μl
Pos 3 WASH
Port 2–4
8 μl
Port
5–7
9
μl
Flow channel diameter
0.8 mm
Valve principle
Motor controlled valve
Functions
Switching 3 positions controlled
from UNICORN
Degree of protection
IP 43
Wetted materials
PEEK (polyetheretherketone)
Chemical resistance
The wetted parts are resistant to organic
solvents and salt buffers commonly used
in chromatography of biomolecules, except
100% ethylacetate, 100% hexane and
100% tetrahydrofuran (THF)
Power requirement
32 V DC ±10% from the system pump
Power consumption
Up to 9 W
UniNet 2 address
0–9
Inlet and outlet tubing
UNF 10–32 2B ”Fingertights” for capillary
tubing 1/16” outer diameter
Dimensions,
H × W × D
135 × 80 × 120 mm
Weight 1.2
kg
EMC Standards
This device meets the requirement of the
EMC Directive 89/336/EEC through the
following harmonized standards:
•
EN 61326 (emission and immunity)
•
EN 55011, GR 2, Class A (emission)
•
This device complies with part 15 of
the FCC rules (emission). Operation is
subject to the following two conditions:
1
The device may not cause harmful
interference.
2
This device must accept any
interference recieved, including
interference that may cause
undesired operation.
Содержание AKTA Valve INV-907
Страница 1: ...Valve INV 907 Instructions GE Healthcare ÄKTA ...
Страница 2: ......
Страница 4: ...4 Valve INV 907 Instructions 56 8101 01 Edition AE ...
Страница 6: ...6 Valve INV 907 Instructions 56 8101 01 Edition AE ...
Страница 8: ...8 Valve INV 907 Instructions 56 8101 01 Edition AE 1 Introduction ...
Страница 18: ...18 Valve INV 907 Instructions 56 8101 01 Edition AE 5 Troubleshooting ...
Страница 22: ...22 Valve INV 907 Instructions 56 8101 01 Edition AE ...
Страница 23: ...23 Valve INV 907 Instructions 56 8101 01 Edition AE ...