Power stability requirement
Table 2-4: Power stability requirement
IEC 61000-4-11
Voltage dips, short
interruptions and
voltage variations
on mains supply
< 5%T (> 95% dip) for
0.5 cycle;
< 5%T (> 95% dip) for
0.5 cycle;
Mains power
quality should be
that of a typical
commercial or
hospital
environment.
40%T (60% dip) for 5
cycles;
40%T (60% dip) for 5
cycles;
70%T (30% dip) for 25
cycles;
70%T (30% dip) for 25
cycles;
< 5%T (>95% dip) for 5
sec
< 5%T (>95% dip) for 5
sec
Содержание 5489329
Страница 5: ...LOGIQ C9 Series Basic Service Manual i 3 5482298 100 English Rev 5 ...
Страница 6: ...i 4 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Страница 7: ...LOGIQ C9 Series Basic Service Manual i 5 5482298 100 English Rev 5 ...
Страница 8: ...i 6 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Страница 9: ...LOGIQ C9 Series Basic Service Manual i 7 5482298 100 English Rev 5 ...
Страница 10: ...i 8 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Страница 11: ...LOGIQ C9 Series Basic Service Manual i 9 5482298 100 English Rev 5 ...
Страница 12: ...i 10 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Страница 13: ...LOGIQ C9 Series Basic Service Manual i 11 5482298 100 English Rev 5 ...
Страница 18: ...i 16 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Страница 52: ...Introduction 1 28 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Страница 76: ...Site Preparations 2 24 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Страница 118: ...System Setup 3 42 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Страница 206: ...Renewal Parts 9 8 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Страница 242: ...Care and Maintenance 10 36 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Страница 247: ...GE ...