Section 10-7 - Electrical Safety Tests
10-25
.
Table 10-24 Test Equipment and Accessory Description
Dale Part number
Accessory Name
Picture
Description
Vendor MODEL
600/100
46-285647P2
CHASSIS CABLE
Black coil cord with extended
length and black grips.
Used on DALE601/601E to measure
earth resistance and enclosure
leakage current. Also used as
reference lead for external
measurement.
Vendor MODEL
600/102
46-285647P4
CHASSIS GROUND PROBE
Black coil cord with needle probe
for testing receptacles and for
tight spaces.
Also referred to as “Saline Probe” or
“Saline Bath Ground Cable”.
Measures earth resistance and
enclosure leakage current. Also used
for grounding saline baths for
isolation testing of probes.
Used on DALE601/601E. This probe
may be substituted for the 600/100
Chassis Cable, and used as a probe
instead of a clamp.
Vendor MODEL
600/200
46-285647P6
ISO/EXTERNAL LEAKAGE
CABLE
Black coiled cord with red grips.
Standard auxiliary cable for external
measurements of leakage current
and voltage gradient between two
surfaces.
Used on DALE601/601E to measure:
• Point-to-Point Leakage Current
• Probe and Transducer Isolation
Current
May only be connected to the female
connector of the Analyzer, labeled
EXTERNAL
Vendor MODEL
600/202
2107545-2
ULTRASOUND LOGIQ FAMILY
ADAPTER
Use during M.A.P. or Isolation test
(see
Using Probe Adapter to Measure
Probe Isolation (Sink) Current
)