Section 6-4 - Monitor and LCD Adjustments
6-13
Adjust the Contrast value
6.) Adjust the Contrast value by pressing button (2) and (3).
Exit the Contrast Menu
7.) Press the button (1) to go back to the
Color
menu.
8.) Press the button (3) 5 times to select
Return
at the lower left end.
9.) Press the button (1) to go back to the
Advanced Menu
.
10.)Press the button (3) 4 times to select
Exit
at the lower left.
11.)Press the button (1) to exit.
6-4-2-1-3
Color Temperature
Range: 9000K / 11000K / 13000K / 15000K
Property: Reddish <-> Bluish.
Factory default: 11000K
Adjustment: Change the value in Advanced Menu according to the procedure below.
Go into the Color Temperature Menu
1.) Press and hold the button (1) for 15 seconds, and the Advanced Menu appears.
2.) Press the button (3) once to select the
Color
menu ([RGB] icon).
3.) Press the button (1) to go into the
Color
menu. The upper left end icon is highlighted in
the menu.
4.) Press the button (3) once to select
Temperature
menu ([K] icon).
5.) Press the button (1) to go into the
Temperature
menu.
Color Temperature value adjustment
6.) Adjust the Color Temperature value by pressing the button (2) and (3).
Exit the Color Temperature Menu
7.) Press the button (1) to go back to the
Color
menu.
8.) Press the button (3) 4 times to select
Return
at the lower left end.
9.) Press the button (1) to go back to the
Advanced Menu
.
10.)Press the button (3) 4 times to select
Exit
at the lower left end.
11.)Press the button (1) to exit.