GE H
EALTHCARE
D
IRECTION
5162630, R
EVISION
3
LOGIQ™ S6 S
ERVICE
M
ANUAL
10-2
Section 10-2 - Why do Periodic Maintenance
Section 10-2
Why do Periodic Maintenance
10-2-1
Keeping Records
It is good business practice that ultrasound facilities maintain records of periodic and corrective
maintenance. The Ultrasound Periodic Maintenance Inspection Certificate provides the customer with
documentation that the ultrasound scanner is maintained on a periodic basis.
A copy of the Ultrasound Periodic Maintenance Inspection Certificate should be kept in the same room
or near the scanner.
10-2-2
Quality Assurance
In order to gain accreditation from organizations such as the American College of Radiology (USA), it
is the customer’s responsibility to have a quality assurance program in place for each scanner. The
program must be directed by a medical physicists, the supervising radiologist/physician or appropriate
designee.
Routine quality control testing must occur regularly. The same tests are performed during each period
so that changes can be monitored over time and effective corrective action can be taken.
Testing results, corrective action and the effects of corrective action must be documented and
maintained on the site.
Your GE service representative can help you with establishing, performing and maintaining records for
a quality assurance program.
Section 10-3
Periodic Maintenance Schedule
10-3-1
How often should PMs be performed?
The Periodic Maintenance Schedule specifies how often your LOGIQ™ S6 should be serviced and what
items need attention. It is important you have your LOGIQ™ S6 serviced as scheduled in order to retain
its high level of safety, dependability and performance.
Your GE Service Representative knows your LOGIQ™ S6 best and can provide competent, efficient
service. Please contact us for further information and to schedule GE Healthcare Ultrasound to perform
this service for you.
The services and intervals shown in the maintenance schedule assumes that you use your LOGIQ™
S6 for an average patient load (10-12 per day) and not used as a primary “mobile unit”.