Output Wiring
The
benchtop
outputs are connected to removable terminal blocks on
the rear panel. Figure 2-2 on page 2-2 shows the location of the
benchtop terminal blocks and Figure 2-8 below shows the wall-mount
connections.
The
Wall-Mount
OptiSonde
input/output terminal blocks are located
inside the front door as shown in Figure 2-6 on page 2-5. Cabling is
brought in through the glands on the bottom of the unit and wired to
the terminal blocks shown in Figure 2-8 below. Terminal block
screws must be torqued between 0.4 - 0.5 Nm.
Note:
Output programming is described in Chapter 4.
Figure 2-8: Wall-Mount Input/Output Terminal Blocks
ANALOG INPUTS
PIN
1. OUTA +
2. OUTA -
3. OUTB +
4. OUTB -
FERRITE
RS232
PIN
1. TX
2. RX
3. RTN
I/O
RELAY
PIN
1. NO
2. NC
3. COMM
A/C POWER
SHIELD
EARTH GROUND CONNECTION
TEMP. SENSOR
PIN
1. RED
2. WHITE
3. GREEN
4. BLACK
5. SHIELD
WIRE COLORS
NOTE: CUSTOMER MUST INSTALL
SUPPLIED FERRITE (P/N 222-028)
OVER THE END OF TEMPERATURE
SENSOR ON LOCAL AND REMOTE.
Содержание 1111H
Страница 1: ...GE Sensing OptiSonde General Eastern Chilled Mirror Hygrometer User s Manual ...
Страница 9: ...Chapter 1 ...
Страница 10: ...Features and Capabilities Introduction 1 1 Electronics Enclosure 1 2 The System 1 3 Sensors 1 4 ...
Страница 15: ...Chapter 2 ...
Страница 33: ...Chapter 3 ...
Страница 43: ...Chapter 4 ...
Страница 44: ...Programming the OptiSonde Introduction 4 1 Programming Technique 4 2 Programmable Functions 4 3 ...
Страница 57: ...Chapter 5 ...
Страница 68: ...Appendix A ...
Страница 73: ...Appendix B ...
Страница 74: ...Humidity Equations and Conversion Chart Introduction B 1 Vapor Pressure B 1 Humidity B 2 ...
Страница 77: ...Humidity Equations and Conversion Chart B 3 October 2007 Figure B 1 Graphical Humidity Conversion Chart ...
Страница 78: ...Appendix C ...
Страница 79: ...Configuring the Serial Interface Wiring to a Personal Computer C 1 ...
Страница 82: ...Appendix D ...
Страница 83: ...Chilled Mirror Sensors Introduction D 1 Depression D 1 Measurement Range D 2 Comparing OptiSonde Sensors D 3 ...
Страница 87: ...Appendix E ...
Страница 88: ...Theory of Operation and Glossary Theory of Operation E 1 The PACER Cycle E 4 Glossary E 5 ...