ZXR-40A
- 10 -
returns to the initial interface, the measurement wiring may be removed.
5. Precautions
1
When measuring the no-load voltage changing transformer, be sure to exit from the
discharge before switching the tapping switch. Do not switch the tapping switch before the
discharge is completed according to the discharge indication.
2
Do not remove the wiring during the measurement.
3
If the charging progress strip and the current display value stand still for a long time, it
indicates that the value of the resistance value to be measured is probably beyond the
measuring scope of the current, so it is not able to reach its set value. At this time, press the
“Exit” button to return and reselect the current to retry.
4
When replacing the fuse tube and the fittings, please use the same model as this instrument.
6. Troubleshooting
Fault phenomenon
Cause analysis
Solution
Remarks
There is no display
after switching on.
1) The power supply
is not switched on.
Switch on the
power supply
Replace with the
fuse tube of the
same model. Do not
substitute it with the
fuse tubes of other
models.
2) The 10A fuse tube
of the instrument is
not properly fitted or
the circuit is open.
Fit or replace the
fuse tube
There is no current
output
1) The equipment to
be tested is
open-circuited
Check the
equipment and
eliminate the faults.
2) The test circuit is
open-circuited.
Check the test
circuit and
eliminate the
open-circuit fault.
7. Transportation and Storage
■ Transportation
When transporting the equipment, it is suggested using the instrument packing wood boxes and
damping materials of our company to avoid any unnecessary damage and loss during the
transportation.
Do not stack the equipment if it is not transported in a wood box. If the instrument packing box