15/52
EN
7.2. CONNECTING DEVICE TO TU
Any devices used with TU must be compatible with the gas for which TU
is intended for.
Always make sure the device is connected first and then used on the
patient.
7.2.1. CONNECTING DEVICE TO TU
Push the probe (L2) into the TU and lock
it in the working position.
7.2.2. DISCONNECTION OF TU
To disconnect the probe (L2) from TU
hold the probe and press the button (B)
into the final position.
7.3. AFTER EVERY USE
Disconnect the medical devices used
from the terminal unit.
8. CLEANING
Remove general contamination with
a soft cloth damped in oil free oxygen
compatible soap water and rinsed with clean water. Disinfection can be
carried out with an alcohol-based solution (spray or wipes).
If other cleaning solutions are used, check that they are not abrasive and
that they are compatible with brass, plastic materials of components, and
gas.
Do not use cleaning solutions containing ammonia!
Do not immerse in water or any liquid.
Do not expose to high temperature.
9. LIFETIME OF THE PRODUCT, MAINTENANCE
AND SERVICING
9.1. BATCH NUMBER AND PRODUCTION DATE
The number stamped on the terminal unit body consists of the following
data: GCE logo XXXXXXX
XXXXXXX: batch number
For example: The number 7366506 indicates a product with the batch
number 7366506.
L2
L2 B
TU - connected