6.1.3. QUICK COUPLER INSTALLATION (QC)
A
B
C2
C1
D
F
H
J
Guide surface
•
Check presence of o-ring (F) in the housing (H).
•
Fit a quick coupler (D) into the housing (H) and tighten manually. Then
tighten the QC (D) of additional 45° using the pipe wrench for tightening
and spanner for keeping the QC (see the special tools info above) in
the correct position - means quick coupler must be positioned by guide
surface down, in horizontal position (as shown on the picture).
•
Fix a cover (recessed C2 or exposed C1) on the installation box (J).
Make sure the CE label is located in right bottom corner.
•
Fix button (B) to the quick coupler (D).
•
Snap the gas specific plugs (A) into the button (B) according to the con-
nected gas type. If the gas is indicated by two different colours keep the
order of colours from left to right according to EN ISO 9170-1 or national
standard, ex. medical air (AIR) - black-white.