www.gastron.com
08
_
09
4.1. Components
3.3. Electrical Specifications (Standard Type)
ITEMS
SPECIFICATION
Input Voltage(Standard)
※ Customer supplied PSU must meet
requirements IEC1010-1 and CE
Marking requirements.
Absolute min:
Nominal:
Absolute max:
Ripple maximum allowed:
18V
24V
31V
1V pk-pk
Wattage
Max. wattage:
Max. current:
4.8W @+24 VDC
200mA @+24 VDC
Analog output Current
0-20mA(500 ohms max load)
All readings ± 0.2mA
Measured-value signal:
4mA(Zero) to 20mA(Full Scale)
Fault:
0-100% LEL:
100-109%LEL:
Over 110% LEL:
Maintenance:
0mA
4mA - 20mA
20mA - 21.4 mA
22mA
3mA
Analog output current ripple & noise max
±20uA
Wiring requirement
Power
CVVS or CVVSB with shield
Analog
CVVS or CVVSB with shield
Cable Connection Length
Analog
2500m
EMC Protection:
Complies with EN50270
No
NAME
DESCRIPTIONS
1
Housing Body
Protects PCB Board built in Sensor and Housing from external environmental change and shock.
2
Housing Cover
Top surface is built with circular glass to enable monitoring of measurement displayed on LCD.
3
MAIN PCB
Amplifies fine outputs generated from Sensor Element to transmit a converted output in 4~20 mA DC
standard. It sends data to display part.
4
Display PCB
Displays data (measurement) sent from Main/ Terminal (Transmitter) PCB on LCD and displays power
status with a power lamp.
5
Power/Signal Terminal
CN8 consists DC18V~31 24 V power supply and DC 4~20 mA standard output connection terminal
(VISO, +24V, mA, GND, ETH).
6
Sensor Terminal
CN4 is Sensor Connection Terminal.
7
Power LED
It is lamp that turns on when power is supplied.
8
Function Key
It is used to enter function setting mode from gas measuring mode by contacting magnet-bar for 2
sec or longer. It is also used to save changed data from function setting mode.
3.4. Environmental Specifications
ITEMS
SPECIFICATION
Operation Temperature
Transmitter
-40 to 80 ℃
Sensor
Refer to Sensor Specification
Storage Temperature
Transmitter
-40 to 80 ℃
Sensor
Refer to Sensor Specification
Operation Humidity
Transmitter
5 to 99% RH (Non-condensing)
Sensor
Refer to Sensor Specification
Pressure Range
90 to 110KPa
Max. air velocity
6m/s
[Figure 2. GTD-2000ExW Components]
GTD-2000ExW
Instruction Manual
4. Name and Description of Each Part