oFF mode is to deactivate relay output and oN mode is to activate output.
-.
Alarm2 Blink (“A3bL”) is displayed
This function is to set that alarm1 relay output is continuously on/off (1 sec
interval) during operation of buzzer
-
Press “Func” key to go into Alarm3 Blink output setting mode.
-.
If pressing “RESET” key, returns to density display mode
-. This mode is to change Alarm3 Blink output type a
nd “on” and “off” is
changed in turns when pressing “▲” or “▼” key.
-.
If desired Alarm3 Blink output type is displayed, press “FUNC” key to set
the type and go to next item.
-.
If pressing “RESET” key, returns to density display mode
-. Alarm3 Blink outpu
t types are “ on” and “ oFF”.
oFF type is to deactivate Alarm3 Blink output and oN type is to activate
Alarm3 Blink output.(this output is activated only when Alarm3 relay output
type is “on”)
-. T
he completion message “End” is displayed for 2 seconds and return to
density display mode.
6-3-6. Maintenance setting
-.
After power “ON”, press “Test” and “Reset” keys at the same time when
“SELF” is flickering and go into maintenance setting mode.
-.
Please make sure to press “Reset” key in maintenance mode, in order to
return to density display mode.
-. This mode is to select functions to set current input and output values, and
“-In-” or “-oUt” are displayed when pressing “▲” or ▼” key.
-.
If desired mode is displayed, press “func” to go into selected mode.
-
.“-In-” mode is to set current input value and “-oUt” mode is to set current
output value
-.
If pressing “RESET” key, returns to density display mode
-. This mode is to select
“-In-” and to set 20mA (input current) and message
is shown as
“IA20”.
-. Press
”Func” key to go into number display mode (input current).
-.
If pressing “RESET” key, returns to density display mode
-. This mode is to display current (20mA) inputted to (mA) terminal.
-. If displayed value is stable and press
”Func” key to set the value and go to
next item.
-.
If pressing “RESET” key, returns to density display mode
-. This mode is to select
“-Out” and to set 4mA, and the message is shown as
“A-04”.
-. Connect Ampere meter to 4~20mA Output terminal.
Press
“▲” or ▼” key to show 4mA in Ampere meter and push “Func” key to
set output current and go to next item.
-.
If pressing “RESET” key, returns to density display mode
-.This mode is to set 20mA output current and message is shown as
“A-20”.
-. Connect Ampere meter to 4~20mA Output terminal.
Press “▲” or ▼” key to show 20mA in Ampere meter and push “Func” key
to set output current and go to next item
-.
If pressing “RESET” key, returns to density display mode