General description
Detector type
Ultrasonic Gas Leak
Detector for fixed
installations
Sensor technology
Stainless steel Micro-
phone technology
Detector frequency range 25 kHz – 70 kHz
Dynamic range
44 dB – 104 dB SPL
Response time
Instant
Detection coverage (leak
rate 0.1 kg/s)
5 - 20 meter radius
Power requirements (minimum)
When using 4-20mA and Relay output
Input Voltage
9 Vdc
Input Current
57 mA
Input Power
600 mW
When using relay output only
Input Voltage
9 Vdc
Input Current
39 mA
Input Power
350 mW
Safety description
Power supply
Ui
28 Vdc
Ii
220 mA
Pi
950 mW
Ci
14 nF
Li
10 µH
output methods
Analogue interface
0 mA
Low supply voltage/No power
3 mA
Unit locally inhibit
4 - 20 mA
44 dB – 104 dB SPL
Relay (option 1 or 2 operator selectable)
Option 1
Alarm relay (switch on preset
trigger level)
Option 2
Error relay
naMur/dry contact (operator selectable)
NAMUR output
Normal (NC)
760
Ω
±1%
Alarm (NO)
10,422k
Ω
±1%
Environmental data
Operating
temperature
- 40 °C to + 75 °C
Humidity
0 to 100 % relative humidity
Ingress Protection
IP66
Atmospheric
conditions
Normal air pressure and
oxygen content
Construction
Housing
Stainless steel AISI 316L
Weight
2.24 kg
Vibration sensitivity
0.13 m/s² perpendicular to the membrane will trig-
ger the unit in the most sensitive range (44dB SPL)
Dimensions
182 x 145 mm
(7.2 x 5.7 inch)
Two gland/conduit
entries
M20 x 1.5
RF emission and immunity
Tested according to:
General
electrical safety
EN61010-1
RF emission
EN61000-6-4
RF immunity
EN61000-6-2
Certifications
ATEX: II 2G EEx ia
IIC T4
IECEX: EEx ia IIC T4
SIL 1/2 rated
15.0 -
Technical daTa
182 mm / 7.2 inch
1
4
5
m
m
/
5
.7
i
n
c
h
Installation and User Guide for Gassonic Surveyor (BB6019-01)