acTion
diSplaY
Switch the Gassonic 1701 on. The version
number will display briefly. You must now se-
lect the detector type.
1
Select
type
BuTTon
Select “Gassonic Surveyor” by pressing
the Down button and then press Enter.
2
Use default filter 25 kHz by pressing the
Enter button.
Select test with the Down button.
3
4
BuTTon
Gassonic
Surveyor
Filter 25 kHz
Select
test
Select gain test by pressing the Enter button.
5
Test
gain
Activate the test by pressing the Enter button.
6
When the test is finished, the Gassonic 1701 beeps twice, and
shows a message, corresponding to the Gassonic Surveyor’s
internal settings. In this example the trigger level is 79 dB and
the rotary switch is in position 5. If the test is more than ± 3 dB
out of tolerance, perform calibration.
If the test was faulty the Gassonic 1701 beeps three times and
a No response message displays on the screen. Press Enter to
try again, or check the “Trouble shooting”-chapter
7
Activate
“TEST”
Meas: 79
Pos: 5
No response
7.0 -
Gain TeST pRoceduRe
Installation and User Guide for Gassonic Surveyor (BB6019-01)