EN - Page 7 

 

Complete cleaning of the Hydrogen Electrode 

Sometimes a complete cleaning of the HydroFlex is required 

 especially when there 

are visible deposits on the electrode. 

 
Remove the hydrogen cartridge.  
Place the hydrogen electrode in 1 mol/l nitric acid for 1 hour. 

 

 

 

 
Then place the entire hydrogen electrode in distilled or 
demineralized water for 24 hours. 

 

 

Dry the hydrogen electrode at about 120 °C (put the 
hydrogen electrode upside down) for 24 hours.  
Now the hydrogen electrode is completely cleaned. 

 

 

 
Install a new hydrogen cartridge and activate it. 

 

 

 

 

 

 

Содержание HydroFlex

Страница 1: ...HydroFlex Wasserstoffreferenzelektrode Hydrogen Reference Electrode Artikelnummer Item number 81000 81010 User Manual Version 3 0...

Страница 2: ...Gaskatel GmbH Lilienthalstrasse 146 D 34121 Kassel 49 0 561 59190 info gaskatel de www gaskatel com Languages DE Bedienungsanleitung Seite DE 1 11 EN Manual Page EN 1 11...

Страница 3: ...Lieferumfang 2 Konformit tserkl rung 2 Allgemeines 2 Zu dieser Anleitung 2 Bestimmungsgem er Gebrauch 2 Sicherheitshinweise 2 Verwendung 3 Ger te berblick 3 Bedienung 3 Betriebsbedingungen 10 Technis...

Страница 4: ...es Produkt erf llt alle f r dieses Produkt g ltigen EU Richtlinien Allgemeines Zu dieser Anleitung Lesen Sie diese Anleitung sorgf ltig Um eine lange Lebensdauer sowie eine zuverl ssige Nutzung zu gew...

Страница 5: ...hfolgenden Hinweise da sonst die Elektrode besch digt oder aber Messergebnisse verf lscht werden k nnen Bei Arbeiten mit Chemikalien sind alle relevanten Sicherheitsbestimmungen des Herstellers und de...

Страница 6: ...ecksilber und silberfrei Bedienung 1 Inbetriebnahme Entfernen Sie vorsichtig die Schutzkappe von der Elektrode Oben am Elektrodenkopf drehen Sie mit einem 3 mm Sechskantschl ssel die Laufzeit im Uhrze...

Страница 7: ...Laufzeit von 6 Monaten erreicht wird muss umgehend die Wasserstoffcartridge ausgetauscht werden siehe 4 Wartung Andernfalls bildet sich ein Unterdruck in der Elektrode so dass Fl ssigkeit in das Inner...

Страница 8: ...die Cartridge vorsichtig heraus Sie kann als Batterie entsorgt werden Nehmen Sie die neue Cartridge zur Hand und fetten Sie zuerst das Gewinde ein z B mit Silikonfett Drehen Sie die Cartridge nun in...

Страница 9: ...Entfernen Sie die Wasserstoffcartridge Stellen Sie die Elektrode f r 1 h in 1 mol l Salpeters ure Dann legen Sie die gesamte Elektrode f r 24 Stunden in destilliertes oder demineralisiertes Wasser Tr...

Страница 10: ...ten manchmal dauert der Ausgleich von Konzentrationen einfach viel l nger als erwartet Gegebenenfalls berpr fen Sie das Potential in einem anderen Elektrolyten wie z B 1 mol l Salzs ure siehe 3 berpr...

Страница 11: ...trode ist verunreinigt und es stellt sich ein Mischpotential ein Kontrollieren Sie die Laufzeit der Wasserstoffquelle Gegebenenfalls muss die Wasserstoffcartridge ausgetauscht werden 4 Wartung Manche...

Страница 12: ...alilauge Natronlauge bis 50 wt Technische Daten Gesamtl nge 120 mm Schaftl nge 80 mm Schaftdurchmesser 8 mm Elektrodenschaft Polytetrafluorethylen PTFE Elektrodenkopf Polytetrafluorethylen PTFE Messel...

Страница 13: ...n aus unserer Produktpalette kostenlos Haftungsausschluss Das Produkt wird kontinuierlich weiterentwickelt Aus diesem Grund ist es m glich dass Teile der Anleitung technische Daten oder Bilder in dies...

Страница 14: ...s Overview 2 Delivery scope 2 Declaration of conformity 2 General 2 About these instructions 2 Intended use 2 Safety iInstructions 2 Use 3 Device overwiew 3 Operation 3 Operating conditions 10 Technic...

Страница 15: ...tion of conformity This product complies with all EU directives applicable to this product General About these instructions Read these instructions carefully To ensure a long service life and reliable...

Страница 16: ...s otherwise the electrode may be damaged or the measurement results may be falsified When working with chemicals all relevant safety regulations of the manufacturer and the laboratory must be respecte...

Страница 17: ...of mercury and silver Operation 1 Commissioning Remove the protective cap from the electrode carefully On top of the electrode head turn the runtime clockwise to position 1 with a 3 mm hex screw driv...

Страница 18: ...running time of 6 months is reached the hydrogen cartridge must be replaced immediately see 4 Maintenance Otherwise a vacuum will form in the electrode and liquid can be soaked 3 Checking the HydroFl...

Страница 19: ...the Hydrogen cell Remove the cartridge carefully It can be disposed of as a battery Take a new cartridge and grease the thread e g with silicone grease Screw the cartridge into the head of the electr...

Страница 20: ...emove the hydrogen cartridge Place the hydrogen electrode in 1 mol l nitric acid for 1 hour Then place the entire hydrogen electrode in distilled or demineralized water for 24 hours Dry the hydrogen e...

Страница 21: ...ated solutions to weak concentrated solutions Await the setting time sometimes the balancing of concentrations lasts longer than expected If necessary check the potential in another electrolyte such a...

Страница 22: ...mits Deposits e g red or brown on the electrode The hydrogen electrode is contaminated and a mixed potential is set Check the runtime of the hydrogen supply If necessary the hydrogen cartridge must be...

Страница 23: ...stic potash caustic soda solution Up to 50 wt Technical data Total length 120 mm Shaft length 80 mm Shaft diameter 8 mm Electrode shaft Polytetrafluorethylene PTFE Electrode head Polytetrafluorethylen...

Страница 24: ...des from our product range free of charge Exclusion of liability The product is subject to continuous further development For this reason it is possible that parts of the instructions technical data o...

Отзывы: