EN - Page 12
Temperature
Many processes take place at higher temperatures. Therefore, it is often necessary to
heat the test cells.
Gaskatel FlexCell test cells can be heated via integrated heating elements. These are
placed into corresponding boreholes and can simply be pulled out before cleaning the
test cells.
Please use our temperature control box 83900 for temperature control, because only a
well-regulated temperature ensures comparable measurement results.
Operation
1.
Assembling the test cell
First, the working electrode may have to be cut to size:
Width 30 mm, length 50 mm
The thickness of the working electrode is almost arbitrary.
The active area in the electrolyte is 3 cm².
The contact screws in the working electrode holder must be turned out a little so that the
electrode cannot touch these contacts at first.
The cut working electrode is placed between the two silicone sealings or EPDM O-rings.
The fixing pins ensure that the electrode cannot slip.
Integrated heating
elements: easy to contact
and can be pulled out if
necessary.