EN - Page 4
Optionally available accessories
Hydrogen reference electrode Mini-HydroFlex 81020
Temperature control box 83900
Analyte compartment PTFE 83110
Analyte compartment PP 83210
Declaration of conformity
This product complies with all EU directives applicable to this product.
General
About these instructions
Read these instructions carefully. To ensure a long service life and reliable use, all
instructions mentioned in this manual must be observed and followed. Please keep these
instructions handy.
You can access the instructions and further information at
Intended use
The FlexCell test cell is suitable for electrochemical measurements on gas diffusion
electrodes and other flat samples such as metals.
We supply appropriately cut silicone seals and O-rings made from EPDM. Make sure that
the seals are well seated before assembling the test cells.
The test cells are designed for the typical three-electrode set-up. A counter electrode is
already integrated.
Both test cells can be heated via integrated heating elements. They may only be operated
with the
Gaskatel temperature control box 83900.
They are connected in parallel to the temperature control box.
The
test cell FlexCell PTFE
may be operated up to
150°C.
The test cell
FlexCell PP
measuring cell may be operated up to
80°C
.
The
filling level of the electrolyte must be checked regularly
, as the level can change
significantly during the measurement.
The measuring cells may only be operated in a laboratory environment.