SGC and SGC Plus User’s Manual

UM SGC v3.0

8 of 15

DETECTOR MAINTENANCE

Bump Test Interval (bUP)

Using the GCT IR Link or GCT Manager, detectors can be programmed to alert the user if a bump test is due. This 

interval can be set anywhere from 1 to 365 days. *

Note the unit default is to have no bump interval programmed

.

If a detector is due for a bump test, the Display will alternate between the months remaining and “bUP”. In addition, 

the detector will emit alternating flashes (left and right) every 5 seconds. And the “test” icon will remain even after a 

button push.

This alert can be cleared by either placing the detector in a Clip Dock or, if the bump interval is set, by manually 

applying gas to the detector. To manually clear the alert, press the Power Button down once and wait for the GAS to 

show on the display while the TEST icon flashes. The detector will wait for 45 seconds for the target gas to be applied, 

or a button press to skip the bump test. If the SGC is bumped while showing GAS, then it will record as a bump test 

in the event log instead of as an exposure. If no gas is applied, it will return to the normal screen and will not record 

anything in the event log.

O₂

 

Detector Calibration

 

GCT recommends users of the O₂ Single Gas Clip to bump test the detector before using each day.

Single Gas Clip Oxygen (O₂) detector factory default will prompt the user to calibrate the detector every 30 days. The 

user will be prompted by the screen flashing CAL, please see calibration instruction below:

Calibration Instructions

 

Only perform O₂ calibration in normal air (20.9% Oxygen) that 

is free of hazardous gases.

1. Press and hold down the yellow button for four (4) seconds.

2. The screen will display CAL and the O

2

 icon will flash in the 

lower left hand side.

3. After a successful calibration, the detector will emit one (1) 

beep, vibration and LED flash.

4. After an unsuccessful calibration, the detector will beep, 

flash, and continue to display calibration. If after a few failed 

calibrations please contact GCT customer support at 877-525-

0808.

Содержание Multi Gas Clip

Страница 1: ...Single Gas Clip Plus Single Gas Clip User s Manual G A S C L I P T E C H C O M...

Страница 2: ...n 5 Turning the Detector On 5 Turning the Detector Off 5 Performing a Self test 5 Alarms 7 Detector Maintenance 8 Bump Test Interval bUP 8 O2 Detector Calibration 8 Cleaning 9 Hibernation SGC Plus Onl...

Страница 3: ...ion The CO and H2S versions of both the Single Gas Clip and SGC Plus do not require calibration for the life of the product The detector should be bump tested before each day s use with a known concen...

Страница 4: ...ur l talonnage Les versions CO et H2S du Single Gas Clip et du SGC Plus ne n cessitent pas d talonnage pendant toute la dur e de vie du produit Le d tecteur doit tre test avant chaque utilisation quot...

Страница 5: ...of 15 DETECTOR COMPONENTS ENTRY DESCRIPTION 1 Audible Alarm Port 2 Display 3 Alarm Bar LEDs 4 Maintenance LEDs 5 Power Button 6 Sensor 7 Infrared IR Communication Window 8 Alligator Clip with Safety...

Страница 6: ...play Details The detector utilizes a special high viewing angle LCD that is designed to enhance the screen visibility In the absence of gas it displays life remaining In those cases where gas is prese...

Страница 7: ...and the sensor is tested Below we outlined the most common screen configurations however if the detector has been programmed via the GCT IR Link or has been exposed to gas additional screens may appea...

Страница 8: ...ctor will be reset Note while the value will be cleared on the display the value will be held in the detector s log See the event log section for further details This value can be cleared on the next...

Страница 9: ...life Please discontinue use Alarm Types LOW ALARM Audible alarm One 1 slow beep every second Visual alarm One 1 slow flash every second Vibrating alarm One 1 slow vibration every second HIGH ALARM an...

Страница 10: ...button press to skip the bump test If the SGC is bumped while showing GAS then it will record as a bump test in the event log instead of as an exposure If no gas is applied it will return to the norma...

Страница 11: ...ve the event log by pressing the save event log before hibernating 4 Click on the Hibernate Button acknowledge the event log message 5 Keep in front of the GCT IR Link until the Hibernate OK message i...

Страница 12: ...of logs you want to download a Event Logs Only b Events Partial Data Log last week of data c Events New Data Logs Only All data logs from last time you pulled the logs to current d Full Logs approxim...

Страница 13: ...CALCAP Replacement calibration cap for the Single Gas Clip or the Single Gas Clip Plus Single Gas Clip Dock P N SGC DOCK portable chargeable all in one docking station in a durable portable case for a...

Страница 14: ...ts Newer events replace older events Warranty Full 2 years SGC or 3 years SGC Plus Gas Range Resolution Accuracy T90 Gases H2S 0 100 ppm 0 1 2 ppm 30 s CO 0 300 ppm 1 5 ppm 30 s O 0 30 vol 0 1 0 7 vol...

Страница 15: ...with a damp cloth Intrinsically Safe S curit lntrins que Class I Division 1 Groups A B C and D T4 Ex ia IIC T4 Ga Class I Zone 0 AEx ia IIC T4 Ga 40 C H2S 40 C CO 35 C O Ta 50 C CAN CSA C22 2 No 60079...

Страница 16: ...arranty length of coverage for products sold without an Activate Before Date shall be two 2 years beginning upon the original end user s documented date of purchase or one 1 year from the product s or...

Страница 17: ...aforementioned information Please note that the defective product must be returned to GCT address which is 305 W FM 1382 Suite 540 Cedar Hill TX 75104 with provided prepaid shipping label GCT will ret...

Отзывы: