Multi Gas Clip Simple User’s Manual
V1.0
4
Portable gas detectors you can count on.
17 of 20
Failures/FAQ’s
•
“E
rr
”
-
If “
E
rr
”
is displaying in place of a particular sensor's Gas Reading
, this
sensor has failed and is
therefore
disabled. Contact GCT for either warranty
replacement or replacement
sensor(s).
Retest function described in this document
can be used to retest a sensor. If
a unit repeatedly fails retest, then the detector
needs to be returned for evaluated for possible sensor replacement.
Under no circumstances should a unit displaying ‘Err’ be used in a potentially
hazardous area
•
“BUMP DUE”
-
If the detector is displaying “
BUMP DUE
”, the detector is either due
for a bump test because of a scheduled test or has failed its last bump test. Refer to
the
Bump Testing
section for more details.
•
“CAL DUE”
-
If the detector is displaying “
CAL DUE
”, the detector is due for a
calibration because of a scheduled interval. Refer to the
Calibration
section for
more details.
•
"LOW BAT"
- If the battery is too low, the detector will display “LOW BAT” and then
turn off. This warning will only display at the detector's end of life.
See the GCT website
,
www.gascliptech.com
,
for further details or contact GCT for pricing and
availability.
•
Replacement Sensors:
o
H
2
S & CO Dual-Tox Sensor
(P/N: MGC-SE-4DT)
o
LEL Sensor - Infrared
(P/N: MGC-S-LEL)
o
O
2
Sensor
(P/N: MGC-SE-O2)
•
Replacement Filters:
o
Filter - Pack of 10
(P/N: MGC-
S-
FILTER-10)
o
Filter - Pack of 50
(P/N: MGC-
S-
FILTER-50)
•
Replacement Detector Components -
Various authorized replacement/spare
parts for
Multi Gas Clip Simple detectors are available