Multi Gas Clip Simple User’s Manual
V1.0
4
Portable gas detectors you can count on.
12 of 20
•
When a bump interval has been programmed into the detector, but a bump
test is not yet due, the current bump test status will show:
“
L. BUMP___
”
(the
date of the last bump test) followed by
“BUMP
XXX
d.
”
(the number of days until the
next scheduled bump test is due). Quickly press the power/menu button while the
display shows
“
BUMP
XXX
d.
”
to force a manual bump test. Audible, visual &
vibrating prompting to apply gas will start. The display will alternate
between “
BUMP
DUE
” and
“
APPLY GAS
”
. Snap the MGC-S Calibration Cap (provided with the
detector) into place over the sensor ports. Apply gas to the detector at a flow rate of
0.5 LPM and do not disturb while test is being performed. Once all of the sensors
have
been tested, the detector will go into alarm. Remove the MGC-S Calibration Cap
and
the detector will return to normal operation after a short period of time. The
bump
due date will be automatically reset based on the new bump date. The bump
test will
be recorded in the Bump Log.
Press power/menu
button
When a bump test interval has not been programmed into the detector, the
current bump test status will show:
“
L. BUMP___
”
(the date of the last bump test).
Quickly press the power/menu button while the display shows
“
L. BUMP___
”
to
force a manual bump test. Audible, visual & vibrating prompting to apply gas will
start. The display will alternate between
“
BUMP XXXd.
”
and
“
APPLY GAS
”
. Snap
the MGC-S Calibration Cap (provided with the detector) into place over the sensor
ports. Apply gas to the detector at a flow rate of 0.5 LPM and do not disturb while
test is being performed. Once all of the sensors have been tested, the detector will
go into alarm. Remove the MGC-S Calibration Cap and the detector will return to
normal operation after a short period of time. The bump test will be recorded in the
Bump Log.
Press power/menu
button
BUMP
____d.