background image

Multi Gas Clip Simple User’s Manual

 

V1.0

4

Portable gas detectors you can count on. 

13 of 20 

When a bump test has failed, the current bump test status will show: 

L. 

BUMP

FAILED

 (indicating that the last bump test had failed). Audible, visual & vibrating

prompting to apply gas will automatically start. The display will alternate between

BUMP DUE

 and 

APPLY GAS

. Snap the MGC-S Calibration Cap (provided with

the detector) into place over the sensor ports. Apply gas to the detector at a flow
rate of 0.5 LPM and do not disturb while test is being performed. Once all of the
sensors have been tested, the detector will go into alarm. Remove the MGC-S
Calibration Cap and the detector will return to normal operation after a short period
of time. The bump test will be recorded in the Bump Log.

If a sensor fails the bump test, display will show 

“E

RROR

 and designate which

sensor 

(CO, H2S, O2 or LEL) 

did not pass. The failed sensor will be disabled and

will need to be replaced (see 

“E

RROR

 in 

Failures/FAQ’s

 section).

During a manual bump test, the audible, visual & vibrating prompting to apply gas 
will continue for approximately two minutes. If gas is not applied within the two 
minutes, then the bump test will automatically fail.  The prompts will cease, the 
detector will return to normal operation, but the display will continually flash 

“BUMP 

DUE”

 until the detector has been successfully bump tested. Please note, to abort a 

bump test at any time, press the power/menu button once and the detector will 
return to normal operation.   

Note - Users may simply apply gas to the detector at any time during normal 
operation to run a bump test without going through the steps above.
 Be 
aware, this will not be recognized by the detector as a Manual Bump Test 
therefore it will be recorded into the Event Log as an Event, not into the 
Bump Log as a Bump Test. 

L

.

Содержание Multi Gas Clip Simple

Страница 1: ...User s Manual G A S C L I P T E C H C O M MultiGasClip Simple ...

Страница 2: ...isplay Details 5 Alarms 7 Default Alarms 7 Alarm Behavior 7 Detector Menus Options 8 Main Menu 8 Status Menu 8 Option Menu 9 Adjustable Options 9 Detector Maintenance 11 Bump Testing 11 Manual Bump Test Instructions 11 Calibration 14 Manual Calibration Instructions 14 Battery 14 Storage 14 Detector Records Logs 15 Event Log 15 Bump Log 15 Calibration Log 15 Data Logs 15 Accessories and Replacement...

Страница 3: ...ire or explosion Keep new and used batteries away from children Caution Before each day s usage sensitivity must be tested on a known concentration of methane equivalent to 25 50 of full scale concentration Accuracy must be within 0 20 of actual Accuracy may be corrected by calibrating the equipment Attention Avant chaque utilisation la sensibilité doit être testée sur une concentration connue de ...

Страница 4: ... Simple User s Manual V1 04 Portable gas detectors you can count on 4 of 20 Detector Components Included in the box Multi Gas Clip Simple Detector MGC S Calibration Cap MGC S Quick Start Guide Calibration Certificate ...

Страница 5: ... of days remaining until end of life During a calibration or bump test the Calibration Test Mode icon 3 will appear and flash indicating it is time to apply gas The Battery Indicator 4 is displayed as a battery icon which is solid as long as the battery is not low The battery won t need charging or replacing during the life of the detector When the battery approaches end of life it will proceed th...

Страница 6: ...old down the power menu button Upon activation the DODUP EDUV will IODVK WKH GHWHFWRU ZLOO YLEUDWH DQG D VWDUW XS WRQH ZLOO VRXQG while the detector self WHVWs the YLVXDO YLEUDWLQJ DQG DXGLEOH DODUPV Gas readings will begin to display immediately Each sensor will show a chasing 0 for the sensor reading while it is stabilizing and being self tested Once all sensors have completed the warm up and st...

Страница 7: ...er various alarm conditions Alarm Condition Audible Alarm Vibration Alarm Visual Alarm Low slow beep slow vibration slow LED flash High fast beep fast vibration fast LED flash TWA slow beep slow vibration slow LED flash STEL fast beep fast vibration fast LED flash Multi slow fast beep slow fast vibration slow fast LED flash Sensor Error fast beep fast vibration fast LED flash Low Battery 1 3 of ba...

Страница 8: ...e Status Menu is accessed by pressing the power menu button once during the SHOW STATUS prompt The detector will display in the following order 1 L CAL ___ the last calibration date 2 CAL DUE or CAL IN XXXd denotes if a calibration is due or how many days until it is due will be displayed only if a calibration interval has been set up 3 L BUMP ___ the last bump test date 4 BUMP DUE or BUMP XXXd de...

Страница 9: ... separate alarm threshold values that tell the detector when to go into alarm Alarm limits may be disabled by setting them to zero Caution Confirm alarm levels with local laws regulations before operation SAFE Display If Safe Display is enabled screen will display SAFE if there are no gas or instrument alerts Self test Lock If a sensor fails a self test the detector shows Err on the display and go...

Страница 10: ...k software Calibration Interval The calibration interval controls how often the detector notifies the user to calibrate the sensors The interval can be individually adjusted for each sensor using the GCT IR Link software Calibration Gas When the detector is calibrated it scales the sensor readings to match the concentrations of the applied gases The calibration gas concentrations can be adjusted t...

Страница 11: ... Life Remaining 2 Current Date Time 3 User Programmed Text Message 4 SHOW STATUS Prompt When the screen displays SHOW STATUS quickly press and release the power menu button to display the current calibration status followed by the current bump test status 1 2 3 4 Press power menu button The next step depends on the following situations When a bump interval has been programmed into the detector and...

Страница 12: ...ion Cap and the detector will return to normal operation after a short period of time The bump due date will be automatically reset based on the new bump date The bump test will be recorded in the Bump Log Press power menu button When a bump test interval has not been programmed into the detector the current bump test status will show L BUMP___ the date of the last bump test Quickly press the powe...

Страница 13: ...t display will show ERROR and designate which sensor CO H2S O2 or LEL did not pass The failed sensor will be disabled and will need to be replaced see ERROR in Failures FAQ s section During a manual bump test the audible visual vibrating prompting to apply gas will continue for approximately two minutes If gas is not applied within the two minutes then the bump test will automatically fail The pro...

Страница 14: ...he calibration parameters Once calibration is complete the detector will display the next calibration date before returning to normal alarming operation If a sensor fails to calibrate the detector will display ERROR an error message for the failed sensor Check your gas connections and concentration before attempting a second calibration If a sensor fails to calibrate after a second attempt contact...

Страница 15: ...test and so on Bump tests are differentiated from normal events when the alarm condition occurs inside of an MGC S Dock or MGC S Wall Mount Dock or when the detector is manually bump tested according to the Bump Testing section The detector records the bump status for each test as follows Date and time of bump test If the bump test was performed manually or with the MGC S Dock Wall Mount Dock Each...

Страница 16: ...ticulate Filter MGC S Calibration Cap 10 ft Sampling Hose and 3 ft Calibration Test Hose with Quick Connect Hand Aspirator Pump Assembly 0 5Lpm regulator and GCT IR Link o MGC S Confined Space Kit w Gas P N MGC S CSK GAS MGC S Confined Space Kit plus 58L Quad Gas 25 ppm H2S 100 ppm CO 18 O2 and 50 LEL o Replacement Air Stone Particulate Filter Pack of 3 P N MGC PF MGC S Hand Aspirator Kit P N MGC ...

Страница 17: ...cause of a scheduled test or has failed its last bump test Refer to the Bump Testing section for more details CAL DUE If the detector is displaying CAL DUE the detector is due for a calibration because of a scheduled interval Refer to the Calibration section for more details LOW BAT If the battery is too low the detector will display LOW BAT and then turn off This warning will only display at the ...

Страница 18: ...68 Warranty Full 2 years Gases Gas Range Resolution Accuracy T90 H2S 0 100 ppm 0 1 2 ppm 30s CO 0 500 ppm 1 5 ppm 30s Combustible 0 100 LEL 0 1 5 LEL 30s O2 0 30 vol 0 1 0 7 vol 15s Sensor Type H2S CO O2 Single plug in electrochemical cell Combustible Soldered infrared IR User Options User Message Language Low Alarm High Alarm STEL Alarm TWA Alarm TWA Method TWA Interval STEL Interval SAFE Mainten...

Страница 19: ...s operated in an industrial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Document name MGC Simple User manual Document number GC006 D01 Date May 1 2018 Revision 1 04 Contact Information Gas Clip Technologies Inc 305 W FM 1382 Suite 540 C...

Страница 20: ...n of unapproved parts on the product The obligations set forth in this warranty are conditional on the following 1 Proper storage installation calibration use maintenance and compliance with the User s Manual instructions and any other applicable recommendations of GCT 2 The buyer promptly notifying GCT of any defect and if required promptly making the product available for correction No goods sha...

Отзывы: