MGC Pump User’s Manual
UM-MGC-P v1.06
17 of 19
DETECTOR SPECIFICATIONS
Size
5.7 x 3.0 x 1.5 in. (147 x 78 x 38.3 mm)
Weight
12.4 oz. (352 g)
Temperature
-4˚ to 122˚F (-20˚ to +50˚C )
Humidity
5% to 100% RH (non-condensing)
Battery Life
IR (infrared) version
52 hours (continuous)
Pellistor version
12 hours (continuous)
Charge Time
4 to 6 hours
Alarms
Visual, Vibrating, Audible (minimum 95 dB)
Pump, Low, High, STEL, TWA, OL (Over Limit)
LEDs
4 Red alarm bar LEDs
Yellow backlight (activated on button press)
Red backlight (activated on alarm condition)
Yellow maintenance notification LED
Display
Alphanumeric Liquid Crystal Display (LCD)
Logs
25 Bump tests
25 Events
25 Calibrations
Continuous 1-second data logging (> 2 months typical capacity)
Tests
Full-function self-test upon activation
Sensors, battery and circuitry tests run continuously
Ingress Protection
IP 67
Warranty
Full 2 years
Gas Inlet
1/8’’ Stainless steel barbed fitting
Pump Draw
Sample from up to 75 ft. (23 m)
Sample Delay Time
Less than 1 second per 1 ft. (0.3m) with 1/8 in. ID tubing
Gases
H
2
S
0 – 100.0 PPM (0.1 PPM increments)
CO
0 – 500 PPM (1 PPM increments)
Combustible (LEL)
0 -100.0% LEL (0.1% increments)
O
2
0 - 30.0% vol. (0.1% increments)
Sensor Type
H
2
S, CO, O
2
: Single plug-in electrochemical cell
Combustible: Plug-in infrared (IR) or pellistor
User Options
User Message, Language, Low Alarm, High Alarm, STEL Alarm, TWA Alarm, TWA Method, TWA
Interval, STEL Interval, SAFE, Maintenance Notification, Self-test Lock, Dock Lock, Pump Test, Sensor
Enable, Calibration Interval, Bump Interval, Calibration Gas