DETECTOR MAINTENANCE

Applying Gas from a Gas Cylinder

The detector contains an internal pump that is continually drawing in gas through the gas inlet. The best way to 
apply gas from a cylinder is through the use of a demand flow (vacuum-actuated) regulator. This type of regulator 
will supply the exact amount of gas that the detector is drawing with the pump. Alternatively, a pressure-based 
regulator rated for at least 300 ml/min can be used. In that case, a T (or Y) fitting should be placed between the 
regulator and the detector to vent the excess gas that exceeds the flow rate being sampled by the detector’s 
pump.

Battery

The battery is displayed as both a percentage and a 3-bar battery icon. The percentage calculation is approximate 
and can be used to provide a rough estimation of the time remaining.

When the detector determines that only 20 minutes of run time remains, it will beep, flash and display “LOW BAT” 
on the screen. This will repeat with 10 minutes remaining, and then continuously for the last 5 minutes of run time. 
When the battery has expired, the detector will give 5 long beeps and flashes, display “OFF” and then shut down.

To charge the detector, either plug it into an AC outlet using the provided charging adapter. You may also charge 
the MGC Pump using an MGC Vehicle Charger or MGC Multi Charger. Both are authorized GCT MGC Pump optional 
accessories. The detector will cycle the battery icons in a charging fashion until the battery is fully charged.  It is 
advisable to turn the detector off before charging.

   

 DO NOT charge the instrument in a combustible atmosphere.

   

 DO NOT charge the instrument in temperatures above or below the specified range of 0°C to 40°C.

   

 DO NOT substitute any other battery type than specified and supplied by GCT.

   

 DO NOT use any charger other than that supplied or offered by GCT.

Storage

Store the detector in a safe, dry place between 32°F and 77°F (0°C - 25°C). For longer storage, the detector battery 
should be discharged to about 40%. After storage, charge the battery and verify the detector calibration before 
use.

   

  Review the warranty period and the “Activate before…” date that is printed on the product box label.

Содержание Infrared

Страница 1: ...User s Manual G A S C L I P T E C H C O M Multi Gas Clip Pump Infrared Multi Gas Clip Pump Pellistor...

Страница 2: ...Detector Records Logs 7 Event Log 7 Bump Log 7 Calibration Log 7 Data Logs 7 How to Retrieve Logs Using the GCT IR Link 8 Detector Options Menus 9 Adjustible Options 9 Main Menu 11 Status Menu 11 Opt...

Страница 3: ...nes lead compounds e g tetraethyl lead sulfur compounds phosphorus halogenated hydrocarbons and aerosols Exposure to poisons may impair the accuracy and or response time of the detector This applies t...

Страница 4: ...imiter des nettoyants base d alcool et d agrumes des silicones des compos s de plomb par exemple plomb t tra thyle compos s soufr s phosphore hydrocarbures halog n s et a rosols L exposition des poiso...

Страница 5: ...ion 1 Power Button 2 Audible Port 3 Display 4 Alarm Bar LEDs 5 Maintenance LEDs 6 IR Communication Window 7 Gas Inlet 8 Filter Window 9 Certification Label 10 Model and Serial Number 11 Gas Exhaust 12...

Страница 6: ...bump test the calibration bottle will be displayed when it is time to apply gas The battery is displayed in 3 bars as well as a percentage The percentage calculation is approximate and can be used to...

Страница 7: ...st LED flash Pump Blocked Fast beep Fast vibration Fast LED flash Sensor Error Fast beep Fast vibration Fast LED flash Low Battery 20 minutes remaining single beep flash and battery icon on 10 minutes...

Страница 8: ...w for activation the detector will display LOW BAT and then turn off again Once activated the detector will begin to display gas readings immediately Each sensor will show a chasing O for the sensor r...

Страница 9: ...the alarm condition occurs inside of the MGC Pump Dock or when the detector is manually bump tested according to the Manual Bump Test Instructions The detector records the bump status for each test Da...

Страница 10: ...r 4 Select what kind of logs you want to download a Event Logs Only b Events Partial Data Log last week of data c Events New Data Logs Only All data logs from last time you pulled the logs to current...

Страница 11: ...The detector s firmware version The TWA alarm limits The STEL alarm limits The LOW alarm limits The HIGH alarm limits Adjustable Options The detector s options are configured using the IR Link and th...

Страница 12: ...ble Individual sensors can be disabled A disabled sensor is completely removed from the detector s display for sensor readings alarm limits and calibrations Caution A disabled sensor will not measure...

Страница 13: ...empty the detector will skip ahead If the startup message does not fit on one screen it will scroll right to left twice Sub Menu Prompts Pressing the button during a sub menu prompt will cause the de...

Страница 14: ...10 minutes remaining and then continuously for the last 5 minutes of run time When the battery has expired the detector will give 5 long beeps and flashes display OFF and then shut down To charge the...

Страница 15: ...e detector will return to normal operation in alarm and the bump due date will be reset If a bump test interval has not been programmed into the detector users may simply apply gas to the detector dur...

Страница 16: ...ternal pump and sensor areas free of water and debris If these filters become blocked then the detector will be unable to continue sampling and the pump will enter into a pump blocked alarm Looking at...

Страница 17: ...e back of the detectors that can clip on to clothing or belts 6 Telescoping Probe P N MGC TSPROBE Metal probe that extends to 6 foot MGC Pump Probe Case P N MGC PUMP TSCASE Holster for MGC Pump with s...

Страница 18: ...nal pump motor Replacement Particulate Filters P N MGC PFILTER 10 and MGC PFILTER 50 replacement particulate hard plastic filters in either a 10 pack or a 50 pack Replacement Hydrophobic Filters P N M...

Страница 19: ...inuous 1 second data logging 2 months typical capacity Tests Full function self test upon activation Sensors battery and circuitry tests run continuously Ingress Protection IP 67 Warranty Full 2 years...

Страница 20: ...ntrins que Class I Division 1 Groups A B C and D T4 Class I Zone 1 AEx d ia IIC T4 Gb 20 Ta 50 C 3 8Vdc nominal ANSI ISA 12 13 01 2000 C22 2 No 152 M1984 UL 916 4th Edition UL 913 8th Edition UL 60079...

Страница 21: ...or use 3 Any damage or defects attributable to repair of the product by any person other than the authorized dealer or installation of unapproved parts on the product The obligations set forth in this...

Отзывы: