Leak detector Vakumatik IVF R34
9.3.
Advice in the event of a fault or if the alarm is triggered
In the event of any fault and if the leak detector's alarm has been activated, the indicator lamp
"Alarm red" (1.2) lights up and the inbuilt buzzer sounds (1.18). Qualified staff or authorised
specialists should be called immediately in order to identify the fault.
The acoustic alarm (1.18) can be temporarily switched off using the alarm switch (1.4) at the
leak detector. The yellow lamp (1.5) indicates a disabled buzzer.
Possible reasons for a fault:
-
a leak in the system, resulting the alarm initiation value is reached
-
pump or pressure switch defective
-
connection lines (also exhaust lines), filter, liquid barrier clogged or icy
-
condensate vessels or water traps are full, liquid barrier closed
-
electrical defect
-
pump fuse defective
Reasons for a leak might be:
-
leaky connections, connecting pipes / tubes, condensate vessel or liquid barrier
-
faulty diaphragm or valves within the pump or the pressure switch
-
connecting fittings on the container (tank)
-
leaky container (tank) or leak protection lining
Initial steps for failure detection:
Perform functional testing pursuant to Section 8.2 (test leak detector set points and leak
tightness of the system).
Restart of the system after failure:
After the malfunction or alarm has been corrected, the leak detector must be started back up
as specified in Section 7.0.
10.
Disposal
Do not dispose of electronic devices in household waste!
Ensure that each component is recycled to the environment.