SYSTEM DESCRIPTION
HB_IDL101_E_V222.doc
13
Gantner Instruments Test & Measurement GmbH
Column 1
The type of variable is indicated by a short designation. Up to 32 variables can be input and parameterized for the
Datalogger IDL 101. The short designations and corresponding channels are:
•
AI : Analog Input Channel
•
AO : Analog Output Channel
•
DI : Digital Input Channel
•
DO : Digital Output Channel
•
AR : Arithmetic Channel
•
SP : Setpoint Channel
•
AL : Alarm Channel
•
CO : Controller Channel
Column
2
A variable indication with max. 20 characters can be given for each sensor variable.
Column
3
The type of sensor connected appears for any Analog Sensor Variables. The sensor can be selected from a database,
which can be expanded by further user-defined sensors. This field remains empty for variables with no sensors
connected.
Column 4
The type of measurement is displayed. For analog and digital sensor variables the following types of measurement are
available:
Analog Sensors
Digital Sensors
Measurement of Voltage - Single ended
Digital Status Recording
Measurement of Voltage - differential
Digital Frequency Measurement
Measurement of Current
Digital Progressive Counter
Measurement of Resistance - 2-Wire Technique
Digital Up/Down Counter
Measurement of Resistance - 3-Wire Technique
Digital Quadrature Counter
Measurement of Resistance - 4-Wire Technique
Interval Counter
Resistance Bridge - 4-Wire Technique
Digital Input Set
Resistance Bridge - 6-Wire Technique
Status Output - Host-Controlled
Potentiometric Measurement
Status Output, Process-Controlled
Thermocouples - Cold Junction Compensation
Pulse-Width Modulated Output
Thermocouples - Differential Input
Digital Output Set
Current Output, Var. Between 0 .. 22 mA
Logic Output
Table 2.1 Types of measurement for analog and digital sensors