IDL 101
INSTALLATION
16
HB_IDL101_E_V222.doc
Gantner Instruments Test & Measurement GmbH
3. INSTALLATION
3.1. Mounting / Fixing
The
Datalogger IDL 101 has a snap-on mounting for installation on standard profile rails 35 mm (1.4 inch) according to
DIN EN 50022.
The four straps on the backside of the module will perform the mounting on the DIN rail. First you push the two straps
on the bottom behind the notch of the DIN rail and then you press the module on the DIN rail until the two straps on
the top snap in.
In order to take the module off the DIN rail slide the module sideward off the rail or in case it is not possible lift the
module slightly so that the straps on the top get off the notch and the module can be taken off easily by pulling it off
the DIN rail.
3.2. Protective System
The
Datalogger IDL 101 has an IP20 protective system. If required by the operating site, the modules are to be
installed e.g. in a splash-proof or presswater-resistant shell, as known from electrical installation engineering.
3.3. Ambient Temperature
The permissible ambient temperature for the
Datalogger IDL 101 in operation ranges from -30°C to +60°C. The
permitted storage temperature is between -40°C and +85°C.