DEVICES - pag. DE 3
Usage
1
Before plugging in the trolley make sure that:
-
the features reported on the yellow label correspond to the local ones (250V,
50/60 Hz);
-
The electrical input of the appliances that are to be powered does not exceed
3A for each single socket (total electrical input 9A);
-
Plug-in cables and stretching cables (if present) must comply with standards;
-
Feeder line, outlet, distribution panel and circuit breaker must be able to
interrupt short circuit power supply (CEI 64-8 appendix D);
-
Phase control, section, power, length of line and grounding cables must
comply with CEI EN 60204-1 standard;
-
The output power supply must be grounded;
-
The trolley must not be plugged in if the power supply cable or the connector
are damaged (in this case the power outlet must be repaired by the technical
assistance service).
2
When the power outlet is not used, the power supply cable must be put back into its
special opening.
Removal
Remove the multiple power outlet only in case of extra-ordinary maintenance or for the
disposal of the trolley.
Maintenance
Refer to what is reported on pag. IN 12.
Decommissioning and disposal
Refer to what is reported on pag. IN 14.