U L S A F E T Y C L A S S I F I C A T I O N I N F O R M A T I O N
1. Protection against electric shock (5.1)
The Game Ready power supply (Ault Model MW128) is considered to be Class I
(grounded) equipment.
The Game Ready power supply (GlobTek Model GTM21097-5012) is considered to be
Class II (ungrounded) equipment.
2. Protection against harmful ingress of water (5.3)
This product provides ordinary protection against ingress of water. The Game Ready
system is considered to be Class I (ungrounded) when connected to the Ault Model
MW128 power supply, and Class II (ungrounded) when connected to the GlobTek Model
GTM21097-5012 power supply.
3. Degree of safety in the presence of flammable anesthetics or oxygen (5.5)
Not suitable for use in an oxygen enriched environment or in the presence of flammable
anesthetics.
4. Mode of operation (5.6)
The equipment operates under continuous operation and continuous operation with
intermittent loading if operating the system continuously without a full load of ice. Users
should allow the system to rest for at least 15 minutes between each 90-minute
treatment.
5. Electromagnetic Interference:
This equipment has been tested and found to comply with the limits for medical devices
in IEC 60601-1-2:1994. These limits are designed to provide reasonable protection against
harmful interference in a typical medical installation. This equipment generates, uses and
can radiate radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to other devices in the vicinity. However,
there is no guarantee that interference will not occur in a particular installation. If this
equipment does cause harmful interference to other devices, which can be determined
by turning the equipment off and on, the user is encouraged to try to correct the interfer-
ence by one or more of the following measures: reorient or relocate the receiving device,
increase the separation between the equipment, connect the equipment into an outlet
on a circuit different from that to which the other device(s) are connected, consult the
manufacturer or field service technician for help.
Medical Equipment
With respect to electrical shock, fire, and mechanical hazards only
in accordance with UL 60601-1, CAN/CSA C22.2 No. 601.1
<61ZJ>
Powered by Direct Current
Type B Applied Parts
Attention: Consult accompanying documents before use
Manufacturer