C A R E   A N D   S T O R A G E

Cleaning

The Game Ready System is a valuable part of your therapy equipment. Please handle the 

Control Unit as you would a laptop computer. When you are finished using the system for 

the day, unplug the Control Unit from the electrical outlet. Disconnect both the Wrap and 

AC adapter from the Control Unit. Empty the water and ice from the ice box. The interior 

of the ice box should be cleaned once every three months. To clean the exterior of the 

Control Unit, simply wipe with a soft cloth rinsed in warm water and a mild disinfectant 

containing any of the following agents: 
 

A. Mild detergent

 

B. Lysol  

 

C. Rubbing alcohol (off the shelf)

To clean the internal fluid circuit of the Control Unit fill the ice box 3-4 inches deep with 

hydrogen peroxide (2%), attach any Wrap and run the unit for 2 minutes.

Caution:

 Leave the lid open when using hydrogen peroxide.

Storage and Transportation 

Storage Temperature Conditions: 1

o

 - 50

o

 C (33

o

 - 122

o

 F).

Relative Humidity: 10% - 95% non-condensing.
Store the Game Ready System in the carrying case or on a shelf, preferably with the lid 

open to allow the interior to dry. 

Caution:

 Do not keep in extreme hot or cold temperatures (above 120°F or below 33°F). 

Avoid leaving in the trunk of a hot or freezing car.

Cleaning and Replacing the Filter

Your Game Ready unit is equipped with a special filter designed to keep dirt and debris 

out of the fluid system. To assure consistent flow of water through the system, the filter 

should be checked and cleaned every three months of regular use. 

Caution:

 If the filter becomes clogged, the Control Unit alarm will beep, the ”low flow“ 

icon will flash, and the unit will go into pause mode. Check to be sure the filter  

is not clogged.
In addition to cleaning the filter every three months, it should be replaced after one year 

of regular usage. Call us toll-free at 1.888.GameReady for a free replacement filter.

To clean or replace the filter, follow these steps:

1. Turn off power and unplug the system from its electrical outlet. 
2. With the Control Unit in the upright position, open  

    the ice box door.
3. Empty any water and ice in the ice box.
4. Locate the filter inside the ice box with your hand,  

    as shown in Fig. A.
5. Press the silver metal button on top of the filter, as shown  

    in Fig. B. You should hear a click as the filter is released.
6. Pull the filter straight out from its housing (Note: The  

    release button is part of the housing and therefore remains  

    attached to the inside of the unit).
7. Clean the filter by rinsing it thoroughly under warm running 

    water. Shake it a few times to clear out any remaining debris.
8. Re-insert the filter in the unit by reaching in and snapping  

    it back into the housing. You should hear it click into place.

Figure A

Figure B

Содержание 500055

Страница 1: ...CONTROL UNIT User s Manual MODEL 500055 GAME READY CONTROL UNIT MODEL 550350 PRE PROGRAMMED CONTROL UNIT...

Страница 2: ...ce edema swelling and pain where cold and compression are indicated It is intended to be used by or on the order of healthcare professionals in hospitals outpatient clinics athletic training settings...

Страница 3: ...2...

Страница 4: ...___________________ 11 GETTING STARTED __________________________________________ 12 ADJUSTING SYSTEM SETTINGS______________________________ 14 OPERATING THE SYSTEM IN THE BAG________________________...

Страница 5: ...NINGS Improper placement or prolonged use of Game Ready could result in tissue damage such as frostbite During the course of Game Ready therapy patients should monitor the skin surrounding the treated...

Страница 6: ...e Game Ready warranty Be careful not to trip over the System s power cords and connector hose Be mindful of the power cords and connector hoses to avoid any risk of strangulation PRECAUTIONS Do not op...

Страница 7: ...omagnetic Interference This equipment has been tested and found to comply with the limits for medical devices in IEC 60601 1 2 1994 These limits are designed to provide reasonable protection against h...

Страница 8: ...upright Fig A or on one side Fig B to fit on a shelf Caution Do not put the unit on the side where the control panel is upside down This will cause the unit to leak water Note The Wrap is comprised of...

Страница 9: ...Note Often when priming a larger Wrap or vest the Control Unit alarm will sound after 45 seconds the low flow icon will flash and the Control Unit will stop This is because there isn t enough water r...

Страница 10: ...the unit is turned on its side place it back in its upright position before opening the ice box lid by pulling up on the latch Press the power button to turn off the unit Unplug the unit from the AC a...

Страница 11: ...not sterile do not place directly against open wounds sores rashes infections or stitches May be applied over clothing or dressing For more information on a specific Wrap see the Use Guide packaged wi...

Страница 12: ...be sure that the air bladder of the Wrap is completely deflated before starting treatment on the new user To deflate the Wrap simply unplug the Wrap from the Control Unit and squeeze the excess air o...

Страница 13: ...roll through the Programs Pre Set Program 1 Pre Set Program 2 Pre set Program 3 30 minutes on with 30 minutes on with 30 minutes on with no pressure low pressure medium pressure 30 minutes in sleep mo...

Страница 14: ...y Temperature knob Start pause Timer on Decrease time Timer setting Increase time Pressure setting Alarm silencer Celsius Fahrenheit PRE SET PROGRAM DISPLAY For Pre Programmed Control Units Only Progr...

Страница 15: ...hanging Pressure During Treatment The pressure cannot be adjusted while the system is in use To adjust the pressure press the pause button change the pressure to the desired mode and then press the st...

Страница 16: ...pty the ice box P O W E R I N G T H E S Y S T E M How to Use the AC Adapter The Game Ready System operates using external power through the AC adapter Connect the AC adapter to the Control Unit and pl...

Страница 17: ...d Replacing the Filter Your Game Ready unit is equipped with a special filter designed to keep dirt and debris out of the fluid system To assure consistent flow of water through the system the filter...

Страница 18: ...box lid open when using hydrogen peroxide to clean the interior of the Wrap If the ice box is sealed shut hydrogen peroxide can form a gas that could build excess pressure inside the unit and potentia...

Страница 19: ...k on Apply the same Wrap more tightly Apply a different Wrap If the problem persists call customer service at 1 888 GameReady 426 3732 Pressure Too High Alarm The Control Unit has detected that the Wr...

Страница 20: ...at 1 888 GameReady 426 3732 Self Test Error The Control Unit has detected an electronic problem at startup This is a Hard Error The unit needs to be turned off and restarted Disconnect the Wrap from t...

Страница 21: ...ce Control unit leaks water when placed on its side 1 If the Control Unit is on its side check to be sure that the control panel is not upside down The unit is not designed to rest this way and doing...

Страница 22: ...oduct shall become CoolSystems property upon receipt The replacement Game Ready Control Unit is warranted under this written warranty and is subject to the same limitations and exclusions for the rema...

Страница 23: ...Latch Latch Cup Ice Box Latch Cup Hook 2x screws for Latch Cup Hook Rubber Foot with adhesive Handle Grip Screw Plug Knob P O W E R S U P P L Y D I S C L A I M E R Mains power quality should be that o...

Отзывы: